Functions
7.5 Setting Relays
Table 7- 4
Relay Option List
Not Used
Power Off
High Flow
Low Low
Flow Alarm
Fault Alarm
Soft Fault
Spacing
Empty
Aeration
Reverse Flow
BatchTot
Pos Total
Neg Total
High Energy
Low Energy
Energy Postot
Energy Negtot
Td
Ts
High CE
Low CE
Assigning functions to Relay 1:
1. From the [Dual Path Flow] menu scroll down and highlight [I/O Data Control].
2. Press <Right Arrow> and scroll down to [Relay Setup].
3. To access the [Relay Setup] option list press <Right Arrow>.
114
Not Active
Power Off alarm occurs when power fails.
Flow rate exceeds high flow setpoint.
Flow rate exceeds or falls below flow setpoint.
Flow rate exceeds or falls below flow set points.
System loses receive signal (all paths in fault).
Fault condition - memory mode active.
Sensor spacing needs adjusting.
Empty pipe alarm.
Aeration percentage exceeds alarm set point.
Flow is in negative direction.
Batch/Sample total advances.
Positive total volume advances 1 digit.
Negative total volume advances 1 digit.
High Energy value relay setpoint.
Low Energy value relay setpoint.
Positive energy totalizer advances by 1 digit.
Negative energy totalizer advances by 1 digit.
Temperature diference relay setpoint.
Supply temperature relay setpoint.
High Chiller Effeciency value relay setpoint.
Low Chiller Effeciency value relay setpoint.
Operating Instructions, 12/2014, A5E03086491-AC
FUE1010 IP65 NEMA 4X