hit counter script
Siemens SITRANS L Series Operating Instructions Manual
Siemens SITRANS L Series Operating Instructions Manual

Siemens SITRANS L Series Operating Instructions Manual

Rotating paddle switch
Hide thumbs Also See for SITRANS L Series:

Advertisement

Quick Links

SITRANS L
Rotating Paddle Switch
SITRANS LPS200
Operating Instructions
7ML5725 (compact)
7ML5726 (extended with protection)
7ML5727 (cable extension)
7ML5728 (angled extension)
7ML5730 (rigid extension)
01/2023
A5E34210883-AC
Introduction
Safety notes
Description
Applications
Installing/mounting
Connection
Service and maintenance
Technical specifications
Dimension drawings
Notes for use in hazardous
locations
Technical reference
Product documentation
and support
1
2
3
4
5
6
7
8
9
A
B
C

Advertisement

Table of Contents
loading

Summary of Contents for Siemens SITRANS L Series

  • Page 1 Introduction Safety notes Description SITRANS L Applications Rotating Paddle Switch SITRANS LPS200 Installing/mounting Connection Operating Instructions Service and maintenance Technical specifications Dimension drawings Notes for use in hazardous locations Technical reference Product documentation and support 7ML5725 (compact) 7ML5726 (extended with protection) 7ML5727 (cable extension) 7ML5728 (angled extension) 7ML5730 (rigid extension)
  • Page 2 Note the following: WARNING Siemens products may only be used for the applications described in the catalog and in the relevant technical documentation. If products and components from other manufacturers are used, these must be recommended or approved by Siemens. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems.
  • Page 3: Table Of Contents

    Table of contents Introduction ............................5 Operating instructions scope....................5 Safety notes ............................7 Safety marking symbols ....................... 7 Description............................. 9 SITRANS LPS200 Overview ....................9 Features ..........................9 Applications ............................11 Installing/mounting ..........................13 Food grade material......................13 Ambient temperature (housing)/process temperature............15 Process cautions ........................
  • Page 4 Table of contents Dimension drawings ..........................33 Dimensions and temperature ranges .................. 33 Extensions ......................... 34 Measuring vanes........................ 35 Notes for use in hazardous locations....................37 Notes for use in hazardous locations .................. 37 Max. surface temperature and temperature code ............... 37 Technical reference..........................
  • Page 5: Introduction

    Introduction Note • Please follow the installation and operating procedures for a quick, trouble-free installation and to ensure the maximum accuracy and reliability of your device. • This manual applies to SITRANS LPS200 only. Operating instructions scope This manual will help you set up your device for optimum performance. SITRANS LPS200 Operating Instructions, 01/2023, A5E34210883-AC...
  • Page 6 Introduction 1.1 Operating instructions scope SITRANS LPS200 Operating Instructions, 01/2023, A5E34210883-AC...
  • Page 7: Safety Notes

    Safety notes Safety marking symbols In man‐ Description product earth (ground) terminal (shield) protective conductor terminal   Dispose of in an environmentally safe manner, and according to local regulations. WARNING: refer to accompanying documents (manual) for details. CAUTION: Observe electrostatic discharge precautions prior to handling electronic components within the wiring compartment.
  • Page 8 Safety notes 2.1 Safety marking symbols SITRANS LPS200 Operating Instructions, 01/2023, A5E34210883-AC...
  • Page 9: Description

    Description Note • Installation, maintenance, and commissioning must be performed by qualified technical personnel. • The device must be used only in the manner outlined in this instruction manual. • This product is intended for use in industrial areas. Operation of this equipment in a residential area may cause interference to several frequency-based communications.
  • Page 10 Description 3.2 Features SITRANS LPS200 Operating Instructions, 01/2023, A5E34210883-AC...
  • Page 11: Applications

    Applications • Grain, feed, cement, plastic granulate, wood chips, rice, soybeans, etc. • Low or high bulk density materials. SITRANS LPS200 Operating Instructions, 01/2023, A5E34210883-AC...
  • Page 12 Applications SITRANS LPS200 Operating Instructions, 01/2023, A5E34210883-AC...
  • Page 13: Installing/Mounting

    Installing/mounting Note • Installation shall be performed by qualified personnel and in accordance with local governing regulations. • For heavy material, only top mounting of paddle switch is recommended. • Compact SITRANS LPS200 is recommended for side mounting on bins for low or intermediate levels.
  • Page 14 Installing/mounting 5.1 Food grade material After mounting and as a precaution against water ingress into the housing, turn the housing clockwise until the cable entries point downward. Note • Ensure measuring vane is out of any region that is susceptible to build-up. •...
  • Page 15: Ambient Temperature (Housing)/Process Temperature

    Installing/mounting 5.3 Process cautions Ambient temperature (housing)/process temperature ‑40 °C (‑40 °F) ambient and process temperature for version with heating of housing ① T amb. ② T process Process cautions CAUTION Keep the device out of the path of falling material. SITRANS LPS200 Operating Instructions, 01/2023, A5E34210883-AC...
  • Page 16: Sensitivity

    Installing/mounting 5.4 Sensitivity CAUTION Protect shaft and measuring vane from falling material. Sensitivity ① Sensitivity spring ② High ③ Medium ④ Universal voltage version shown. SITRANS LPS200 Operating Instructions, 01/2023, A5E34210883-AC...
  • Page 17 Installing/mounting 5.4 Sensitivity To adjust the sensitivity of the device, use pliers to move the sensitivity spring to the appropriate position (factory setting is medium). The optional measuring vanes can also improve sensitivity. SITRANS LPS200 Operating Instructions, 01/2023, A5E34210883-AC...
  • Page 18 Installing/mounting 5.4 Sensitivity SITRANS LPS200 Operating Instructions, 01/2023, A5E34210883-AC...
  • Page 19: Connection

    Connection WARNING Opening device • Open device only when supply voltage is switched off. Note • All field wiring must have insulation suitable for at least 250 V AC. • A disconnect switch shall be in close proximity to the equipment and within easy reach of the operator.
  • Page 20: Universal Voltage, Dpdt, Fail-Safe

    Connection 6.1 Universal voltage, DPDT, fail-safe Note • Before opening the lid, ensure there are no dust deposits around the device. Ensure that the atmosphere around the device is settled. • Use a fuse for the signal output and power supply (max. 10 A, fast or slow, HBC, 250 V). •...
  • Page 21: Ac Or Dc Version, Spdt

    Connection 6.2 AC or DC version, SPDT AC or DC version, SPDT ① Signal output: Micro switch, SPDT contact Max. 250 V AC, 5 A, non-inductive Max. 30 V DC, 4 A, non-inductive External fuse: max. 10 A, fast or slow, HBC, 250 V ②...
  • Page 22: Switching Logic (Universal Voltage, Dpdt)

    Connection 6.3 Switching logic (universal voltage, DPDT) Switching logic (universal voltage, DPDT) FSH: Use this setting when sensor is used as a full detector. Power failure or line break is regarded as "full" signal (protection against overfilling).   FSL: Use this setting when sensor is used as an emp‐ ty detector.
  • Page 23: Signal Output: Delay (Universal Voltage, Dpdt)

    Connection 6.4 Signal output: delay (universal voltage, DPDT) ① ② Signal output: delay (universal voltage, DPDT) ① Sensor covered → free Factory setting = 3 seconds ② Sensor free → covered Factory setting = 1 seconds SITRANS LPS200 Operating Instructions, 01/2023, A5E34210883-AC...
  • Page 24: Alarm Output - Fail-Safe Option (Universal Voltage, Dpdt)

    Connection 6.5 Alarm output - fail-safe option (universal voltage, DPDT) Alarm output - fail-safe option (universal voltage, DPDT) ① Yellow or green ② ③ Fault ④ No fault Switching and timing behavior: If the vane is not covered, the rotating vane shaft will send pulses at 20 second intervals.
  • Page 25: Switching Logic (Ac Or Dc Version, Spdt)

    Connection 6.6 Switching logic (AC or DC version, SPDT) Switching logic (AC or DC version, SPDT) SITRANS LPS200 Operating Instructions, 01/2023, A5E34210883-AC...
  • Page 26 Connection 6.6 Switching logic (AC or DC version, SPDT) SITRANS LPS200 Operating Instructions, 01/2023, A5E34210883-AC...
  • Page 27: Service And Maintenance

    Please note the following: • The user is responsible for all changes and repairs made to the device. • All new components must be provided by Siemens. • Restrict repair to faulty components only. • Do not re-use faulty components.
  • Page 28 Service and maintenance 7.2 Unit repair and excluded liability SITRANS LPS200 Operating Instructions, 01/2023, A5E34210883-AC...
  • Page 29: Technical Specifications

    Technical specifications Note Siemens makes every attempt to ensure the accuracy of these specifications but reserves the right to change them at any time. Power   • 115 V AC, 50 to 60 Hz, 4 VA • 230 V AC, 50 to 60 Hz, 4 VA •...
  • Page 30: Alarm Output

    Technical specifications 8.4 Mechanical • Hinged measuring vane • 35 g/l (2.2 lb/ft ) when vane is covered by 10 cm (4 inch) of material • 70 g/l (4.4 lb/ft ) when vane is covered by more than 10 cm (4 inch) of material •...
  • Page 31: Environmental

    Technical specifications 8.6 Process -Angled extended 5.2 kg (11.4 lb) -Cable 4.0 kg (8.8 lb) Environmental Indoor/outdoor • Location Max. 2 000 m (6 562 ft) • Altitude ‑25 ... +50 °C (‑13 ... +122 °F) • Ambient temperature 0 ... 100 %, suitable for outdoor (ingress protection: • Relative humidity Type 4X/NEMA 4X/IP66) • Overvoltage category • Pollution degree Note With the heated enclosure option, the ambient temperature rating improves to ‑40 ... +50 °C...
  • Page 32: Approvals

    Technical specifications 8.7 Approvals Note The device construction allows over-pressure between ‑0.2 and 0.1 bar in hazardous areas. WARNING This product is designated as a Pressure Accessory per Directive 2014/68/EU and is not intended for use as a safety device. WARNING Materials of construction are chosen based on their chemical compatibility (or inertness) for general purposes.
  • Page 33: Dimension Drawings

    Dimension drawings Dimensions and temperature ranges Standard Model Compact version Extended Angle extension Rope option                                                 Dimensions in mm (inch) ①...
  • Page 34: Extensions

    Dimension drawings 9.2 Extensions High temperature model Compact version Extended Angle extension Rope option                       ① ① ① Process temperature ‑25 ... +600 °C L = length 125 ... 300 mm Rope fixing (‑13 ... +1 112 °F) (4.92 ... 8.46 inch) ② 100 mm 3.94 inch 2 000 mm ②...
  • Page 35: Measuring Vanes

    Dimension drawings 9.3 Measuring vanes High temperature model: temperature extended shaft Process temperature 150/250 °C (302/482 °F) 200 mm (7.87 inch) 350 °C (662 °F) 350 mm (11.8 inch) 600 °C (1 112 °F) 400 mm (15.7 inch) Measuring vanes Standard Hinged Rectangular Dimensions in mm (inch) ① Fold together to lead into mounting hole ②...
  • Page 36 Dimension drawings 9.3 Measuring vanes   Vane completely covered Bulk material covers the   Rectangular vane options with bulk material vane up to 10 cm (3.93 inch)   Spring Adjustment Spring Adjustment     Vane Light Central Light Central (factory set‐ (factory set‐...
  • Page 37: Notes For Use In Hazardous Locations

    Notes for use in hazardous locations Notes for use in hazardous locations Process pressure Devices with Ex Approval are approved for atmospheric pressure. A detailed explanation is given below for ATEX and applies analogously to other Ex approvals. The scope of the ATEX directive is generally limited to atmospheric pressure, see ATEX directive 2014/34/EU chapter 1, Article 2 (4) Atmospheric pressure is defined as absolute pressure 0.8bar, see ATEX guideline 50 and IEC...
  • Page 38 Notes for use in hazardous locations A.2 Max. surface temperature and temperature code Enclosure directly mounted to the process connection Max. ambient Max. process tem‐ Max. surface tem‐ Max. surface temper‐ Temperature temperature perature perature (EPL Db) ature (EPL Da) class (division system) 30°C (86°F) 50°C (122°F)
  • Page 39: Technical Reference

    Technical reference Principle of operation A low revolution gearing motor drives a rotating measuring vane. When material reaches the rotating vane, rotation stops, actuating a switch. When the vane is no longer covered, rotation resumes and the switch reverts to its normal condition. The synchronous induction motor is freely suspended within the housing.
  • Page 40 Technical reference B.1 Principle of operation SITRANS LPS200 Operating Instructions, 01/2023, A5E34210883-AC...
  • Page 41: Product Documentation And Support

    – "Download as html5, only PC": Open or save the manual in the HTML5 view on your PC You can also find manuals with the Mobile app at Industry Online Support (https:// support.industry.siemens.com/cs/ww/en/sc/2067). Download the app to your mobile device and scan the device QR code. Product documentation by serial number Using the PIA Life Cycle Portal, you can access the serial number-specific product information including technical specifications, spare parts, calibration data, or factory certificates.
  • Page 42: Technical Support

    (http:// www.siemens.com/automation/csi/service).   Service & support on the Internet In addition to our technical support, Siemens offers comprehensive online services at service & support (http://www.siemens.com/automation/serviceandsupport). Contact If you have further questions about the device, contact your local Siemens representative at Personal Contact (http://www.automation.siemens.com/partner).
  • Page 43: Index

    Index Catalog catalog sheets, 41 Certificates, 41 Customer Support, (Refer to Technical support) Downloads, 41 Hotline, (Refer to Support request) Manuals, 41 Service, 42 Service and support Internet, 42 Support, 42 Support request, 42 Technical support, 42 partner, 42 personal contact, 42 SITRANS LPS200 Operating Instructions, 01/2023, A5E34210883-AC...
  • Page 44 Index SITRANS LPS200 Operating Instructions, 01/2023, A5E34210883-AC...

This manual is also suitable for:

Sitrans lps2007ml57257ml57267ml57277ml57287ml5730

Table of Contents