hit counter script

Siemens SITRANS FC310 Compact Operating Instructions page 25

Coriolis flowmeters
Hide thumbs Also See for SITRANS FC310:
Table of Contents

Advertisement

Orienting the sensor
The sensor operates in any orientation. The optimal orientation depends on the process fluid
and the process conditions. Siemens recommends orienting the sensor in one of the following
ways:
1. Vertical installation with an upwards flow (self-draining)
Figure 3-3
2. Horizontal installation, tubes down (recommended for liquid applications)
Figure 3-4
3. Horizontal installation, tubes up (recommended for gas applications)
Figure 3-5
Note
Hygienic applications
In EHEDG certified applications the flowmeter must be installed vertically as shown in 1
above.
SITRANS FC310
Compact Operating Instructions, 01/2019, A5E44932112-AB
Vertical orientation, upwards flow
Horizontal orientation, tubes down
Horizontal orientation; tubes up
(EHEDG certificate in preparation)
Installing/mounting
3.2 Installation instructions
25