Test Equipment; Procedure; Test Menu Utility Softkey; Description - HP 8340B Manual

Synthesized sweepers
Hide thumbs Also See for 8340B:
Table of Contents

Advertisement

Chapter
16.
Test Menu Utility Softkey
16.0 Description
The
test
menu Utility softkey provides
you with
the capability of changing the synthesizer's
defined option configuration (option configuration defined by the Cal Constants) from your
computer, displaying data
from previously run tests, or displaying the test status
of previously
tested instruments- The feature which allows
you to
change the synthesizer's
defined option
configuration should be used after an option has been
added
to
or
deleted from the
synthe¬
sizer.
16.1
Test Equipment
None Required
16.2 Procedure
1.
After the Utility softkey provided
In
the test menu has been selected, the program
will
display the following softkey choices:
Chg Option
-
Allows you to
change
the
synthesizer's
defined option configuration
After selecting
this
softkey,
Input
each of the synthesizer's
3
digit
option strings (one at
a
time) and press ENTER after each entry
Press ENTER when finished
inputting the options and
then check
the
option
list.
Upon completion of the option entry, the program
will
automatically update
the calibration
constants
in
the synthesizer and
the data file on the data disc to reflect the option configuration you
entered
Dsp Data
-
Allows you to display or print the test data from previously run tests
The data displayed
will
be for the
instrument
whose
serial
number
is
currently
in
the computer's
memory and for
the type of testing you
selected while
in
the
TEST SELECTION
menu (I.e
,
either the Opera¬
tion Verification or the
Attenuator Calibration). After
selecting this
softkey, the program
will
display the test menu and
allow you to
output
all
the test data (ALL
DATA softkey), output the data
for one
test
(SELECT
softkey), or return to the
UTILITIES
menu
(EXIT
softkey) The
SELECT
softkey outputs
the
data for
the test
at
which
the cursor points. Use the keyboards UP or
DOWN arrow
keys or the
rotary control knob
on the
keyboard to
position the cursor.
Prt
Status
-
Allows you to display
or
print the test status for the synthesizer
whose
serial number
is
currently
in
the
computer's memory. The
status displayed
will
be for the type of testing
you
selected while
in
the
TEST SELECTION menu
(i.e., either the
Operation Verification or
the Attenuator Calibration menu).
Exit
-
Returns
you to the test menu-
Select
the desired
softkey and respond
to the
prompts provided by the program.
HP 8340A/B and HP 8341A/B
Test Menu
Utility
Softkey