If the ZERO calibration fails for one or more sensors, the detector will give a
single beep and a single flash and both the Test Pass icon
icon
will blink for 5 seconds.
If the ZERO calibration fails for all sensors, the detector will give a single beep
and a single flash and only the Test Fail icon
all Pass
note
If any sensor has failed, repeat the ZERO calibration ensuring that the
detector is in fresh air. If a second failure occurs, contact Honeywell
Analytics or their agent.
Warning
under proper calibration procedures, repetitive calibration failures could
indicate that the sensor is either approaching its end of life, or it has been
seriously contaminated, or both.

4-3. Span Calibration (for Flammable and Toxic Sensors Only)

Calibrate the detector at least every 6 months (CH
exposure to contaminants. User can perform the SPAN calibration with  gases at the
same time or with a single gas. When SPAN calibration is performed with a single gas,
the detector detects the supplying gas automatically.
To carry out the SPAN calibration, the user needs the following accessories, which
are available from Honeywell Analytics (See Section  Optional Accessories):
Calibration gas cylinder of known concentration (multi-gas mix) as follows:
Gas
CH
4
CO
H
S

• A flow regulator supplying the gas at 00 mL/min flow rate
• Tubing for use between the regulator and the flow adaptor
X4 Series
some Pass/Fail
Recommended
Calibration Gas
Concentration
50% LEL
50 ppm
5 ppm
will blink for 5 seconds.
, CO, H
S), depending on use and
4

Concentration Range
Calibration Gas
0 ~ 50% LEL
50 ~ 00 ppm
0 ~ 50 ppm
and Test Fail
all Fail
5