EN - english Instructions for installation and operation ® METPOINT Measurement accessories for the measurement in systems up to 40 bar...
Page 2
Dear customer, ® Thank you for deciding in favour of the METPOINT OCV measuring device with sampling probe. Please read these installation- and operating instructions carefully before mounting and starting up the ® ® METPOINT OCV measuring device, and follow our directions. Perfect functioning of the METPOINT OCV, and thus exact measurement and monitoring of the residual oil content in compressed air, can only be guaranteed when the provisions and notes stipulated here are strictly adhered to.
Operating conditions Operating conditions ® 2.1 Operating conditions METPOINT Max. permissible operating overpressure PS: 16 bar g Working pressure compressed air: 7 bar g Min. / max. permissible operating temperature of the sensor unit: +5° C / +55° C Min. / max. permissible operating temperature of the evaluation unit: +5°...
BEKO TECHNOLOGIES GMBH. This symbol stands for actions which can also be implemented by the operators, provided that they are skilled and authorised accordingly. Measurement accessories METPOINT® OCV 40 bar...
Safety instructions Safety instructions In addition to these operating instructions, it is imperative to adhere to the relevant accident prevention regulations, the generally accepted safety rules, the EU directives and country-specific standards and provisions. Safety labelling Safety instructions whose non-observance represents a threat to life and physical condition or can lead to property damage are labelled in accordance with DIN 4844 and ANSI Z535.
Safety instructions General safety instructions DANGER! Compressed gas! Risk of serious injury or death through contact with quickly or suddenly escaping compressed gas or through bursting plant components. It is the obligation of the operator to ensure that the connected pressure generator is protected against the exceeding of the maximum operating overpressure and of the ®...
Safety instructions In the event of fire, do not use water to extinguish it (neither in the vicinity of the system nor directed directly at the system). Special advice for plants under pressure in accordance with the 97/23/EC Pressure Equipment Directive The proper use of the measuring device is the basic requirement for safe operation.
Safety instructions Special safety instructions IMPORTANT! Compressed air! Exceeding the maximum pressure can lead to damaging the plant. Observe the indications regarding the maximum pressure on the type plate! IMPORTANT! Operational reliability at risk! Incorrect installation can compromise the operational reliability and affect service measures. Do not exceed the permissible volume flow and working pressure! Do not operate below the working pressure! NOTE!
Field of application and proper use of the METPOINT® OCV 40 bar measurement accessories ® Field of application and proper use of the METPOINT OCV 40 bar measurement accessories ® The METPOINT OCV measurement accessories serve to expand the operating pressure range of the ®...
Technical data Measurement accessories METPOINT® OCV Name for the measurement in systems with an operating pressure of up to 40 bar BEKO TECHNOLOGIES GMBH Germany, Manufacturer 41468 Neuss, Im Taubental 7 Compressed air, free from aggressive, corrosive, Medium caustic, toxic, flammable, and fire-accelerating...
Transport and set-up Transport and set-up IMPORTANT! Danger through damaged components! Under no circumstances should you start up damaged components. Defective components can impair the operational reliability and cause further damage. ® The function and service life of the METPOINT OCV in the version for the measurement in systems with an operating pressure of 40 bar depends on the conditions at the place of installation.
Operating principle of the METPOINT® OCV 40 bar ® Operating principle of the METPOINT OCV 40 bar ® The METPOINT OCV in the version for an operating pressure of 40 bar was developed for the detection of vaporous and gaseous hydrocarbons in compressed air that is free from aggressive, caustic, toxic, flammable, and fire-accelerating constituents.
Description structure and components Description structure and components ® ® Sensor unit METPOINT OCV (component part METPOINT OCV) Pressure reducer (measurement accessories) Connection between the pressure reducer and the ball valve, 1m stainless steel pipe 6 x 1 mm (measurement accessories) Ball valve 3/8“...
Details pressure reducer 10 Details pressure reducer The pressure reducer is intended for a maximum primary pressure of 40 bar(g). The secondary pressure is adjusted to 7 bar(g) and can be read at the manometer. No further settings are required on the part of the customer. Measurement accessories METPOINT®...
Installation 11 Installation 11.1 Mechanical installation ® The installation of the METPOINT OCV and of the 3/8“ ball valve is carried out in accordance with the ® specifications in the separate instructions for the installation and operation of the METPOINT OCV.
Removal from service 13 Removal from service DANGER! High pressure! Risk of serious injury or death through contact with quickly or suddenly escaping compressed air or through bursting plant components. Unpressurise the plant prior to any maintenance or repair works! DANGER! Supply voltage! Contact with non-insulated parts carrying supply voltage involves the risk of an electric...
® For the maintenance of the METPOINT OCV, the specifications in the separate installation and operating instructions apply. Use original BEKO spare parts only. 14.1 Care NOTE! Aggressive cleaning agents and solvents must not be used! 14.2 Calibration ®...