Scale parameters and functions
8.10 DR 10 load cell parameters
Variable
Note
Load cell char‐
Characteristic value of the load cell
acteristic value
(n) [mV/V], the mean value is used if
(Page 80)
there is more than one cell.
1)
Reserve
Rated load of a
Rated load of a load cell
load cell
(Page 80)
1)
Reserve
Reserve
Reserve
Reserve
Reserve
Reserve
Reserve
Reserve
Reserve
Reserve
Parameter for calculation of calibration points with theoretical calibration
1)
8.10.1
Selection of the sampling rate
For better suppression of interferences, set the measuring rate and the line frequency as
follows:
Measuring rate
100 Hz
120 Hz
Interferences can be caused by the power supply grid.
For fast measurements, you can select a measuring rate of 600 Hz (parameter code 2) as of
firmware V1.2.0.
The setting has an effect on the frequency response of the signal path.
The following diagrams show the frequency response:
78
Type
L
REAL
4
REAL
4
REAL
4
REAL
4
REAL
4
INT
2
UINT
2
REAL
4
Line frequency
50 Hz (parameter code 0)
60 Hz (parameter code 1)
Operating Instructions, 08/2019, A5E33715669A-AD
Rw
Default Min.
rw
2
> 0.1
rw
0
rw
60
-
rw
0
-
rw
0
-
rw
0
-
rw
0
-
rw
0
-
Max.
10
-
-
-
-
-
-
SIWAREX WP321