Siemens FIDAMAT 6 Operating Instructions Manual page 118

Gas analyzer for determination of total hydrocarbon content
Table of Contents

Advertisement

Functions
6.6 Configuration
With softkey 2, you set whether the cross-interference correction is to apply only to certain
measuring ranges.
Under
Correction of cross-interference with variable influence of interfering gas
Figure 6-59
Here, a variable influence of interfering gas is active, which must be measured using a
suitable separate gas analyzer and then supplied as an analog current or digital signal (via
ELAN) to the device for correction of cross-interference.
Press softkey 1 again to perform the cross-interference correction via the RS 485 (ELAN)
serial interface.
Here, the same inputs are to be made as for the correction of cross-interference
analog input"
Also enter the channel number and component number of the interfering gas analyzer. The
type of gas assigned to the channel and component as well as the device status appear in
the display.
Note
On the device supplying the correction data, the parameter
(function 73) must be set to
With softkey 2, you set whether the cross-interference correction is to apply only to certain
measuring ranges.
See also
ELAN configuration (function 73) (Page 106)
116
"Cross-interference deflection"
Variable influence of interfering gas
.
"On"
, enter the value of the zero offset.
.
Operating Instructions, 03/2018, A5E00222135-03
"...via
"Measured value telegram"