hit counter script

General Safety Instructions; Safety-Instrumented System - Siemens ULTRAMAT 6 Safety Manual

Gas analyzer for the determination of ir-absorbing gases
Hide thumbs Also See for ULTRAMAT 6:
Table of Contents

Advertisement

Siemens AG
2

General Safety Instructions

2.1

Safety-instrumented system

Definition: Safety-instrumented system
A safety instrumented system executes the safety functions that are required
to achieve or maintain a safe status in a system. It consists of a sensor, logic
unit/ control system and final controlling element.
Example:
A safety instrumented system is made up of an analyzer (for example CO
concentration), a PLC and a control valve (Figure 2-1).
Definition: Safety function
Defined function executed by a safety instrumented system with the
objective of achieving or maintaining a safe system taking into account a
defined dangerous occurrence.
Example:
CO concentration
Definition: Dangerous failure
Failure with the potential to bring the safety instrumented system into a
dangerous or non-functional status.
Description
The analyzer logic unit/control system is a final controlling element combine
to form a safety-instrumented system, which executes a safety function.
Note
This document deals with the ULTRAMAT 6 exclusively as part of a safety function.
SIL Safety Manual ULTRAMAT 6 Gas analyzers
A5E00729183-02
6 of 19

Advertisement

Table of Contents
loading

Table of Contents