Page 2
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
Unimed 1100 (Leakage current measurement only required in the case of a deviation of the local line voltage from the system delivery sta- tus). ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 6 of 46 02.05 CS SD 24 Medical Solutions...
Note on the log book The log book is located in the monitor cart behind the keyboard. It is accessible if the cover on the back of the monitor cart is removed. Siemens AG SPR2-330.815.01.01.02 ARCADIS Avantic Page 7 of 46 Medical Solutions 02.05...
It must also be ensured by the customer that connection to an unsuitable power supply is prevented (e.g., by identification or design measures). As concluding measures, perform and document protective conductor measurements (in Germany, e.g., in the medical device book). ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 8 of 46 02.05 CS SD 24 Medical Solutions...
Before the line voltage is adapted, the power plug on the UPS in the monitor cart must be pulled out, since the monitor cart is not free of voltage after the power plug is pulled out. Siemens AG SPR2-330.815.01.01.02 ARCADIS Avantic Page 9 of 46 Medical Solutions 02.05...
50%. The system must be repaired if the limit is exceeded. Document the measured value in the equivalent leakage current protocol. ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 10 of 46 02.05 CS SD 24...
Disregarding safety precautions can result in death or serious bodily injury. Before work is done on the monitor cart, the power plugs must be pulled out of the UPS. Siemens AG SPR2-330.815.01.01.02 ARCADIS Avantic Page 11 of 46 Medical Solutions 02.05...
Page 12
NOTE Laser emissions! There is no direct hazard to the eye (blinking reflex). Nevertheless do not look directly into the laser beam. ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 12 of 46 02.05 CS SD 24 Medical Solutions...
Compare the measured line voltage with the line voltage label on the back of the moni- tor cart. The measured voltage must agree with the voltage read off from the label. Siemens AG SPR2-330.815.01.01.02 ARCADIS Avantic Page 13 of 46 Medical Solutions 02.05...
Adapt the voltage by changing the connections to transformers T1 and T2. Refer to the ARCADIS Avantic wiring diagram. Fig. 1: Monitor cart line voltage Pos. 1 Power plug Pos. 2 T1 and T2 ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 14 of 46 02.05 CS SD 24 Medical Solutions...
Document the measured values in the "Leakage current test protocol", (Leakage cur- rent measurement / p. 39). Measuring arrangements for system leakage current with tester VDE 0751 Octo- ber 1990 (e.g. Bender tester) Siemens AG SPR2-330.815.01.01.02 ARCADIS Avantic Page 15 of 46 Medical Solutions 02.05...
Page 16
For safety reasons, place the system on an insulated base. • Perform measurement and repeat measurements. • Document the measured values in the "Leakage current test protocol", (Leakage cur- rent measurement / p. 39). ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 16 of 46 02.05 CS SD 24 Medical Solutions...
"The emergency stop switch has been released." • Document the function of the EMERGENCY STOP switch in the "Test protocol, C-arm movement and emergency stop" (C-arm movement and emergency stop / p. 38). Siemens AG SPR2-330.815.01.01.02 ARCADIS Avantic Page 17 of 46 Medical Solutions 02.05...
Semi-transparent slot and iris diaphragm in fluoroscopy procedure. • Collimation of the semi-transparent slot and iris diaphragm using fluoroscopy proce- dure. • Mechanical functions of the monitor cart ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 18 of 46 02.05 CS SD 24 Medical Solutions...
10 A) • Test procedure: Measure between all touchable, conductive parts of the system and the protective conductor wire at the ARCADIS Avantic power plug. Siemens AG SPR2-330.815.01.01.02 ARCADIS Avantic Page 19 of 46 Medical Solutions 02.05 CS SD 24...
• Document the function of the C-arm movements, brakes and warning signals in the test protocol, "C-arm movement and emergency stop" . ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 20 of 46 02.05 CS SD 24 Medical Solutions...
Please clarify in advance with the customer whether the following system configura- tions are required. Service login: Fig. 5: Service login • Select <Options> <Customer service> <Local service>. Siemens AG SPR2-330.815.01.01.02 ARCADIS Avantic Page 21 of 46 Medical Solutions 02.05 CS SD 24...
Page 22
Specific System Settings Fig. 6: Authentication_ • Enter the 6-character password and click OK (see system folder for password). ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 22 of 46 02.05 CS SD 24 Medical Solutions...
Specific System Settings Customer address configuration Fig. 7: Configuration_ • Select <Configuration>. Siemens AG SPR2-330.815.01.01.02 ARCADIS Avantic Page 23 of 46 Medical Solutions 02.05 CS SD 24...
Page 24
Specific System Settings Fig. 8: Configuration • Click <Next>. ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 24 of 46 02.05 CS SD 24 Medical Solutions...
Page 25
Specific System Settings Fig. 9: Local Host • Click <Site Info>. Siemens AG SPR2-330.815.01.01.02 ARCADIS Avantic Page 25 of 46 Medical Solutions 02.05 CS SD 24...
The marked customer-specific data may be adapted. Do not change "Serial No.". • Click <Save>. • Click <Home>. Main system configuration • In the service software, select "Main System". ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 26 of 46 02.05 CS SD 24 Medical Solutions...
Page 27
Specific System Settings Fig. 11: Main_system • Click <Main system>. Fig. 12: Mainsys start • Select ARCADIS Avantic - I.I. type and click <Next>. Siemens AG SPR2-330.815.01.01.02 ARCADIS Avantic Page 27 of 46 Medical Solutions 02.05 CS SD 24...
Page 28
- Buzzer on: Buzzer sound always during radiation. The country-specific regulations must be observed for the max. fluoroscopy time and radi- ation blockage. Factory setting: ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 28 of 46 02.05 CS SD 24 Medical Solutions...
NOTE A maximum tube current limit is required - as far as known - only in Denmark. • Start service program and log in as described under "Service Login". Siemens AG SPR2-330.815.01.01.02 ARCADIS Avantic Page 29 of 46 Medical Solutions 02.05...
Page 30
Specific System Settings Fig. 14: Main_system • Click <Main system>. Fig. 15: Mainsys start • Select ARCADIS Avantic - I.I. type and click <Next>. ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 30 of 46 02.05 CS SD 24 Medical Solutions...
Page 31
Specific System Settings Fig. 16: Current limitation • Click <Current limitation>. • Make country-specific changes (only Denmark) and click <Save>. • Click <Home>. Siemens AG SPR2-330.815.01.01.02 ARCADIS Avantic Page 31 of 46 Medical Solutions 02.05 CS SD 24...
Select "Backup & Restore" in the Service Software Home Menu. • Place the backup CD (located in the monitor cart service compartment) in the CD-ROM drive. Fig. 17: Backup & Restore_Command ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 32 of 46 02.05 CS SD 24 Medical Solutions...
Page 33
Fig. 18: Backup & Restore_Command_Backup • Select the respective drive under Drives. - Save to CD-ROM (standard save). Fig. 19: Backup SW - Settings02 • Select "SW-Settings02" under Packages. Siemens AG SPR2-330.815.01.01.02 ARCADIS Avantic Page 33 of 46 Medical Solutions 02.05 CS SD 24...
Page 34
Exiting backup • Close the window via <Home>. • Remove the backup CD-ROM from the CD-ROM drive. • Store the CD-ROM with the other system documentation. ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 34 of 46 02.05 CS SD 24 Medical Solutions...
Page 35
It is not possible to perform a backup of the database (patient im- ages)! Store the patient images in an archive. Saving to CD-ROM as a long-term archive is prohibited. Siemens AG SPR2-330.815.01.01.02 ARCADIS Avantic Page 35 of 46 Medical Solutions 02.05...
Plug in the USB cable. • Print a test page after the automatic test. • Follow the instructions on the CD-ROM from the manufacturer in the event of problems. ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 36 of 46 02.05 CS SD 24 Medical Solutions...
Configuring the network (optional) and remote service Configure the network according to the Configuration Guide, SPR2-330.843.02.01.xx.xx and the remote connection according to Installation, Siemens Remote Service, SP00-000.816.02.01.xx.xx. Completing protocols After completing all adjustment work steps and check measurements, the responsible technician must sign and date the test protocols and country-specific protocols to confirm that all values have been correctly determined and recorded.
C-arm movement and emergency stop protocol Movement Name Date Initials Yes/No Releasing/locking the brakes Downward movement, position Downward movement, position The warning signals sound as described. Function, EMERGENCY_STOP_button ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 38 of 46 02.05 CS SD 24 Medical Solutions...
General: Local AE title for: HIS/RIS Configuration/DICOM/General Study transfer * Configuration/DICOM/general Print * Configuration/DICOM/general Comments Network data for HIS/RIS network partner (only with HIS/RIS option) Siemens AG SPR2-330.815.01.01.02 ARCADIS Avantic Page 41 of 46 Medical Solutions 02.05 CS SD 24...
Page 43
Network data for DICOM camera in the network Information for Any camera No camera available Device: Version: Manufacturer: Default camera yes Host properties: Host name * Configuration/DICOM/Print devices Siemens AG SPR2-330.815.01.01.02 ARCADIS Avantic Page 43 of 46 Medical Solutions 02.05 CS SD 24...
Page 44
TCP/IP address * Configuration/DICOM/Print devices General node properties: Logical name * Configuration/DICOM/Print devices Application entity: AE title * Configuration/DICOM/Print devices Port number * Configuration/DICOM/Print devices Comment: ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 44 of 46 02.05 CS SD 24 Medical Solutions...
Page 45
Changes to Previous Version 6Changes to Previous Version n.a. / first version Siemens AG SPR2-330.815.01.01.02 ARCADIS Avantic Page 45 of 46 Medical Solutions 02.05 CS SD 24...
Page 46
Changes to Previous Version ARCADIS Avantic SPR2-330.815.01.01.02 Siemens AG Page 46 of 46 02.05 CS SD 24 Medical Solutions...