Table of Contents WARNINGS AND PRECAAUTIONS........1-1 Symbol Descriptions..............1-2 FUNCTIONAL DESCRIPTION ..........2-1 Introduction................2-1 2.1.1 Principle ................2-1 2.1.2 Main features:..............2-2 2.1.3 Comparison with Other Methods of Breast Examination.. 2-3 2.1.4 Included in this Package ........... 2-4 System Overview............... 2-5 Block Diagrams (1 and 2) ............
Page 6
Full Software Installations ............4-1 4.2.3 Installation Procedure............4-2 4.2.4 Formatting the Hard Drive..........4-2 Upgrade of TS2000 Software............. 4-3 Additional Software Installation ..........4-3 4.4.1 WIN NT4 Installation (Including Service Pack 4).... 4-3 4.4.2 Installation of VGA Driver ..........4-5 4.4.3 Installation of Printer Driver ..........
Page 7
RD7409290-0 Rev. F TS2000 Service Manual TROUBLESHOOTING ............6-1 Error Messages ................6-1 6.1.1 Displayed Error Messages ..........6-1 6.1.2 TransScan TS 2000 Error Message ........6-2 6.1.3 Windows NT Error Messages .......... 6-3 Input/Output Problems..............6-3 Port Setup................... 6-5 PM and PSB TESTS ..............
Page 8
Tables of Figures Figure 2-1: Complete TS2000 System in Cart (Front View) ........2-5 Figure 2-2: TS2000 System Block Diagram............2-6 Figure 2-3: System Block Diagram ................2-6 Figure 2-4 Processing Module-Block Diagram............2-9 Figure 2-5 Probe Control Unit Block Diagram ............2-9 Figure 2-6: Reference Signal Generator Block Diagram .........2-10...
Page 9
ONLY by qualified service personnel. In order to prevent electrical shock; do not re move covers or panels. This equipment should be used ONLY in accordance with instructions for use CAUTION: In the event of defective functioning of TS2000, contact the relevant Siemens service department. SPB7-520.840.01.01.02...
TS2000 Service Manual RD7409290-0 Rev. F 1.1 Symbol Descriptions Symbols Description Type BF equipmen Dangerous voltage Attention: Consult accompanying documents ο ο Off: Power disconnection from the mains On: Power connection to the mains NOTE: The CE 0344 mark for this device can be found on the left side of the central processor.
TS2000 image. TS2000’s unique potential to distinguish and detect early signs of cancer is based upon detecting these electrical changes non-invasively. TS2000 (Trans-Spectral Impedance Scanning) is a unique system designed specifically for non-invasive, accurate detection of early stage breast cancer.
TS2000 is safe! The T-Scan method of detection is non-invasive, uses no radiation and complies with electrical safety standards. The system uses low voltages and currents for measurement, which cause no...
16,000 clinical examinations, there were no reports of harm or adverse side affects to the patient or the operator. TS2000 uses a totally different method of detecting breast disease than mammography, combining the results of T-Scan with those of mammography may increase the overall sensitivity and/or specificity.
TS2000 Service Manual RD7409290-0 Rev. F 2.2 System Overview The TS2000 is housed in a sturdy, portable storage cart and comprised of the following elements (refer to Figure 2-1): • Monitor • Main System Console: Computer Base; Processing Module (PM), •...
TS2000 Service Manual RD7409290-0 Rev. F 2.3 Block Diagrams (1 and 2) Figure 2-2: TS2000 System Block Diagram Figure 2-3: System Block Diagram SPB7-520.840.01.01.02...
2.6 Keyboard and Mouse The keyboard and mouse are standard PC compatible devices that are used by the TS2000 for system operation and data entry. 2.7 Main System Console The Main System Console is comprised of the Computer Base (PC), the Processing Module (PM) and the Probe Selection Board (P.S.B.).
TS2000 Service Manual RD7409290-0 Rev. F 2.9 CD-ROM Drive The CD-ROM is used to load software to the system. 2.10 Modem (TeleService Package—Up to Serial # 2087) A modem is included in all units up to serial number 2087. Use the modem for “Remote-Diagnostic” of the system, through a standard telephone line.
TS2000 Service Manual RD7409290-0 Rev. F Figure 2-4 Processing Module -Block Diagram 2.14 Probe Control Unit The Probe Control Unit consists of: • • RAM configurable as 32 KB × 16 or 128 KB × 16 • Flash ROM of 128 KB × 8 •...
TS2000 Service Manual RD7409290-0 Rev. F Acquisition and interrupt handling software is loaded to the RAM before execution, in order to attain maximum performance. 2.15 Reference Signal Generator The Reference Signal Generator (RSG) consists of the following blocks: • A Programmable Sine Wave Generator •...
TS2000 Service Manual RD7409290-0 Rev. F 2.17 Scan Probes (Small Probe, Large Probe) The Scan Probes are housed in a hand-held casing. The probes are connected by a flexible multi-cable to the connection panel on the PSB (Probe Selection Board) located at the back of the system. The mouse...
RD7409290-0 Rev. F The transmitter cylinder is made of stainless steel. 2.20 Software Overview The TS2000 software consists of three major modules. These modules are the main application, the PM software, and the FoxPro database management software. The main application runs on the Computer under MS-Windows NT and controls the entire system.
Page 23
TS2000 Service Manual RD7409290-0 Rev. F Figure 2-7: TS2000 Software Block Diagram SPB7-520.840.01.01.02 2-13...
Page 24
TS2000 Service Manual RD7409290-0 Rev. F SPB7-520.840.01.01.02 2-14...
3. INSTALLATION OF OPTIONS 3.1 System Setup To set up the system, follow the instructions in the System Manual, in the section: “Installation and Start Up.” NOTE: If the AutoRun feature fails to start upon installation of an auxiliary software program, proceed as follows: Click Start|Run.
TS2000 Service Manual RD7409290-0 Rev. F Slide the printer back into the shelf until it is secured. Slide the paper tray out and fill it with “A4” paper. Return the tray to its place. Turn on the system. Turn on the printer.
TS2000 Service Manual RD7409290-0 Rev. F Close Control Panel and turn system off. Install the ISDN modem and connect it to the power supply and the system. NOTE: Insure that the ISDN modem supply voltage is compatible to the previously installed modem.
TS2000 Service Manual RD7409290-0 Rev. F Close “Control Panel” and restart the system. 3.3.4 Verifying Operation Call the telephone number that the Modem is connected When the Modem responds, the customary Modem sound will be heard. 3.4 Internal Zip Drive (Atapi Model) 3.4.1 Hardware Installation...
Printer, ATI software for display adapter, Application & Service CD and one CD with all the necessary software required to operate the system (including TS2000 system software). 4.1 Verifying the Actual Software Version From the Welcome Screen, start a new exam.
TS2000 Service Manual RD7409290-0 Rev. F 4.2.3 Installation Procedure NOTE: If you cannot set the CD-ROM drive as the boot drive, use the Windows NT boot diskettes, that are part of the Windows NT package, to install Windows NT. If you are missing boot diskette from the setof three diskettes,...
4.3 Upgrade of TS2000 Software On a time-to-time basis, TransScan will supply a new version of software for the TS2000 System. The upgrade will be supplied on a CD-ROM, and will be performed by the service engineer. To perform the upgrade: Install the CD with the new software version into the CD- ROM Drive (the CD has an auto-run program).
Page 32
TS2000 Service Manual RD7409290-0 Rev. F To install a fresh copy of Windows NT, press "N," then press "Enter." Leave the disk size as is; press "Enter" to install the software into the directory "\WinNT," then press "Enter" twice. Follow the onscreen instructions.
TS2000 Service Manual RD7409290-0 Rev. F 4.4.2 Installation of VGA Driver NOTE: ATI often changes the wording in their installation procedures. Therefore, there may be slight differences between the text in the following procedure and the actual wording on screen.
TS2000 Service Manual RD7409290-0 Rev. F Select model “Lexmark Optra K 1220 (or another model according to the printer that you have in your system.)from the list ,click “Add” then click “Next” On drivers screen for Lexmark Optra k 1220 PS2 set the following: Select “Port LPT1”, in “Change setting”...
4.4.6 Installation of ODBC Software Insert the CD with the new software version into the CD- ROM Drive Wait 15 seconds; the TS2000 installation program will start. In the welcome dialog box, click “Cancel." In the Exit Setup dialog box, click “Exit Setup."...
TS2000 Service Manual RD7409290-0 Rev. F 1. Run the PreSetup program from the CD. 2. Select “Install ODBC”, and then “Continue”. 3. From the Menu, select “Microsoft Fox Pro Driver” and click “OK”. 4. When finished, click “Close”/ “OK”/ and then close “ODBC”...
TS2000 Service Manual RD7409290-0 Rev. F 4.4.10 Software Installation The software of the P.M. board will be occasionally upgraded and will be supplied on the same CD-ROM as the system software. From the “Maintenance” menu in the exam screen, select “PM”.
Page 38
TS2000 Service Manual RD7409290-0 Rev. F SPB7-520.840.01.01.02 4-10...
5. PARTS REPLACEMENT NOTE: The replacement of the parts described in this section must be followed by a functional test. See section 7.4. Each intervention (replacement or repair) must be recorded in the “records of service visits” in the manual. 5.1 Replacing the Computer Turn the system power off.
Page 40
TS2000 Service Manual RD7409290-0 Rev. F Unscrew the eight screws that secure the computer to the bottom of the main frame. Slide the computer out carefully and slide the new computer into place. Replace and tighten the eight screws that secure the computer in place (at the bottom).
TS2000 Service Manual RD7409290-0 Rev. F 5.2 Replacing the PM Board Turn off the power system. Disconnect all cables from the main console and remove it from the shelf. Open the upper cover of the main system enclosure (Main Console) by unscrewing the four screws (two on each side).
TS2000 Service Manual RD7409290-0 Rev. F 5.3 Replacing the PSB Board (Mounted) 1. Turn off the power system. 2. Disconnect the probes from the PSB. 3. Unscrew the four screws securing the PSB asembly to the TSP. 4. Disconnect the data cable from the PSB.
TS2000 Service Manual RD7409290-0 Rev. F 5.4.2 Jumpers on the I/O Isolation Card The jumpers are located on the card at the base addresses of the serial COM ports and the parallel LPT port. If the card is installed in a PC...
TS2000 Service Manual RD7409290-0 Rev. F 5.7 Spare Parts List NOTE: The following list also appears on the TransScan Service Manual CD- ROM. Description Catalog No. Siemens Catalog No. Cart AY74010357 47 78 945 RVB75 Transmitter Cable WA7408733-0 47 78 960 RVB75...
Page 45
TS2000 Service Manual RD7409290-0 Rev. F Scan Probe 256 AY74410016 47 78 812 RVB75 Probe Selection Board AY74010412 70 41 093 RVB75 Mounted External Zip Drive MD10078 47 78 846 RVB75 Internal Zip Drive MD10089 47 78 838 RVB75 Modular case for TSP...
Page 46
TS2000 Service Manual RD7409290-0 Rev. F SPB7-520.840.01.01.02...
6. TROUBLESHOOTING NOTE: 1) The effectiveness of any repair and part replacement must be verified by a functional test. Refer to Section 7.4. 2) Each intervention (replacement or repair) must be recorded in the “record of service visits” section of the manual. 3) Replace “blown”...
TS2000 Service Manual RD7409290-0 Rev. F Messages Solution • Make sure the program was activated. Parallel card has no number • Check power to the modem Modem Error • Check the connection between the A message on system startup modem and the computer.
TS2000 Service Manual RD7409290-0 Rev. F 6.1.3 Windows NT Error Messages Exit to “Windows”. Click on Start and Select “Programs”/ “Administrative Tools”/ “Event Viewer." Double click on the event to find event details. 6.2 Input/Output Problems Problem Solution • Check the Power cable connections.
Page 50
TS2000 Service Manual RD7409290-0 Rev. F Problem Solution No image on screen during 1. Check serial communication (see “Scan in Target” or section: PM and PSB test). “Screening” modes. 2. Check PM output. • Try adjusting the window level of the The image on the monitor is too dark.
TS2000 Service Manual RD7409290-0 Rev. F Problem Solution • Check the Printer cable connections. The system is not printing clearly or is not printing at • Check the software Printer Driver setup all. • Check for errors on the printer’s front panel and act accordingly.
TS2000 Service Manual RD7409290-0 Rev. F 6.4 PM and PSB TESTS 6.4.1 Preparation Exit system to Windows. Access the Windows NT Explorer. Double-click PM Test in the TransScan directory. 6.4.2 Serial Communication Test Perform Preparation 6.4.1. Click PC2PM to check serial communication between PM and computer.
TS2000 Service Manual RD7409290-0 Rev. F 6.4.4 Probe I.D. Test Connect the small probe to J1, and select ENA in the lower Control section. Click Read; the address 1011 appears in the Data section (b7-b0). Connect the large probe to J1, and select ENA in the control section.
TS2000 Service Manual RD7409290-0 Rev. F In the “Menu Bar," the “Tests” menu will appear. Click on the Stop/ Start button on the small probe/ large probe; at the bottom of the screen, the message “FPS=X.XX appears. NOTE: X.XX must be 8.00-+1.00 for the small probe.
TS2000 Service Manual RD7409290-0 Rev. F Fuses on the PM board are: Function Fuse Value For +5v For –12V (and also –5V) 3/4A For +12V 6.4.10 PSB Board On the PSB, there are 10 red LEDs, indicating the following: Indication...
Page 56
TS2000 Service Manual RD7409290-0 Rev. F The Fuses on the PCB board are: Function Fuse Value +5V main fuse to the board +5V to port 1 +5V to port 2 +5V to port 3 -5V to port 1 -5V to port 2 -5V to port 3 SPB7-520.840.01.01.02...
7. MAINTENANCE NOTE: This section describes the maintenance procedures that the service technician may perform. In general the system does not require any calibration. Phantom Check The service engineer should perform a routine check of the system every time he visits the site, and when installing a system. In his routine visit, the service engineer will check all the cables, connection and general outlook of the system.
TS2000 Service Manual RD7409290-0 Rev. F Functional Tests Perform the Functional Tests as follows: Perform an examination in Scan Mode with the large probe. Perform an examination in Targeted Mode with the large probe. Perform an examination in Scan Mode with the small probe.
8. APPENDIX 8.1 Detailed Hardware Platform Specifications Environmental Operating Temperature 15°C to 40°C Storage Temperature -20° C to 55°C Humidity 40-60% Non-Condensing Probe Shock and Vibration Portable, rugged System System Computer Pentium II A300 MHz (or higher) based Electrical Operating Voltage 115/230-+10% VAC 50/60 Hz-+2HZ Operating Power 250W...
Page 60
TS2000 Service Manual RD7409290-0 Rev. F Functional Measurement Frequency Range 58Hz to 4 kHz Database entries Limited by hard disk space Display Levels 240 gray level ≤ 5 seconds Database Access Time SPB7-520.840.01.01.02...
TS2000 Service Manual RD7409290-0 Rev. F 8.2 Figures AC Out 2 Connectors 13-14 Power Selection Jumpers 115V Out 220V Out PS Power and Computer Power Connectors 11-12 Pinsulated Output Power Selection T1 Trafo Isolated Power 115V Out 220V Out Connectors 7-10...
TS2000 Service Manual RD7409290-0 Rev. F Parallel Cable Serial COM (RS-232) Mains Setup T2 PM AC PS-DC for T1 Main Transformer Figure 8.2.2: Overhead view NOTE: The change of operating voltage of the system must be carried out in the factory only and not in the field.
TS2000 Service Manual RD7409290-0 Rev. F FUSE 0.5A TM7408787-0 AY7408776-0 MD10062 COMMUNICATION CABLE TM7408783-0 230V AC To P.C 230V AC 230V AC To MONITOR 230V AC To MODEM TRANSFORMER AC IN ET00139 4A FUSE 4A FUSE WT00249 VENT Figure 8.2.3: TSP Power Connection 230 VAC...
6A FUSE VENT Date Name Signature DRAWN 6/9/00 Winer TransScan CHECK Title DESIGN TS2000 TSP Power Connection - 115V AC 6/9/00 Winer Size Document Number APPROVED RD7408802-1 APP. QA Date: Monday, September 04, 2000 Sheet Figure 8.2.4: TSP Power Connection 115VAC...
Page 65
TS2000 Service Manual RD7409290-0 Rev. F PM to PCB Power Out to Cable Monitor Power Out to Modem Transmitter Cable Power In Mouse and Keyboard Connections Modem Cable Probe 1 Printer Connection Probe 2 Parallel Cable Main Power Switch Figure 8.2.5: Cable Locations...