Digital Multimeter
PASAR® 323
Instruction Manual
CAT III 600V
ACER AB
Box 261
SE-433 25 PARTILLE
Sweden
Telefon +46 (0)31-446500
Telefax +46 (0)31-444410
www.acer.se
©
Copyright
2005 Acer AB. All rights reserved
1

Advertisement

Table of Contents
loading

Summary of Contents for Acer PASAR 323

  • Page 1 Digital Multimeter PASAR® 323 Instruction Manual CAT III 600V ACER AB Box 261 SE-433 25 PARTILLE Sweden Telefon +46 (0)31-446500 Telefax +46 (0)31-444410 www.acer.se © Copyright 2005 Acer AB. All rights reserved...
  • Page 2: Table Of Contents

    Contents 1.0 Introduction ..............15 2.0 Safety Measures ..............16 3.0 Operation Elements and Connections ......17 4.0 Use of the Instrument 4.1 Preparation and Safety Measures ........18 4.2 Voltage Measurement AC/DC ..........18 4.3 Current Measurment 10A/15A AC/DC ........18 4.4 Current Measurment mA AC/DC ...........
  • Page 3: Introduction

    1.0 Introduction The PASAR 323 is universally usable. The multimeter is built and tested according to the latest safety regulations. It is a valuable help in the manual and industrial area as well as for the hobby electronics technician at all standard measure tasks..
  • Page 4 Prior to using the instrument the user is kindly requested to read the instruction manual and comply with it in all sections. Failure to read the instruction manual or to comply with the warnings and references contained herein can cause danger to life or serious injury and damage to the instrument.
  • Page 5: Operation Elements And Connections

    3.0 Operation Elements and Connections Digital Display Symbol ”-” for negativ voltage Low battery indication Measurement range selection switch Input socket for voltage, resistance, diode and current mA Input socket COM Input socket for current A...
  • Page 6: Use Of The Instrument

    Prior to usage check the instrument for perfect functioning on a known voltage source (see also DIN VDE 0105, Part 1). PASAR 323 has an Auto Range function that will automaticly search for the best measurement range. 4.2 Voltage Measurement AC/DC Do not connect the probes to more than 600V AC/DC.
  • Page 7: Current Measurment Ma Ac/Dc

    The value measured is shown on the digital display. If the measurment value is >10A an acoustic signal is audible. After working with 15A for 30 s, PASAR 323 needs a ”cool down time” of 5 min. 4.4 Current Measurement AC/DC mA Do not carry out current measurements in systems with more than 250V.
  • Page 8: Continuity Test

    If resistance is <60 Ω an acoustic signal is audible. 5.0 Auto-Power-Off Approximatly 15 min after the latest key operation, PASAR 323 switches off automatically. To switch the instrument on again, turn the rotary switch from the OFF position to the desirered function.
  • Page 9: Fuse Replacement

    Replace the case cover and retighten the screws. Now the measurings can be carried out as normal. Please help care for our environment, do not throw the used battery in the normal household dustbin, make sure it is disposed of as hazardous waste, or collected for recycling.
  • Page 10: Technical Data

    7.0 Technical data 7.1 General Display 3½ digit, LC Display, 3999 digits Overload indication left ”OL” is displayed Measurement rate 3 measurements per second Low battery indication Battery symbol at approx. 2.5V Auto Power Off after approx. 15 min Overvoltage catagory CAT III 600V Pollution degree Working temperature...
  • Page 11: Current Ma

    7.5 Current mA Range Resolution Accuracy Overload protection 40mA 10µA ±(1,5%+3 digits) 250V AC/DC 400mA 100µA Burden voltage approx. 10mV/mA Acoustic signal: >400mA 7.6 Resistance Range Resolution Accuracy Overload protection 400 Ω 0.1 Ω 1 Ω 4 kΩ 10 Ω 40 kΩ...