Honeywell AreaRAE Pro Quick Start Manual
Hide thumbs Also See for AreaRAE Pro:

Advertisement

AreaRAE Pro & AreaRAE Plus
QuickStart Guide
RAE Systems by Honeywell
3775 N. First St.
San Jose, CA 95134-1708 USA
Phone: 408-952-8200
Fax: 408-952-8480
Email: customerserv@raesystems.com
www.raesystems.com
Charging The AreaRAE Pro/Plus
Always fully charge the battery before use.
1. Remove the cover over the AreaRAE Plus/Pro's
power port.
2. Plug the AC/DC adapter's barrel plug into the
AreaRAE Pro/Plus' power port.
3. Plug the AC/DC adapter into an AC outlet.
IMPORTANT!
Keep all ports covered when
they are not in use.
The AreaRAE Pro/Plus User's Guide must be carefully
read by all individuals who have or will have the
responsibility of using, maintaining, or servicing this
product. The product will perform as designed only if it is
used, maintained, and serviced in accordance with the
manufacturer's instructions.
Never operate the monitor when the rear cover is re-
moved. Remove rear cover, sensors, and/or battery only
in an area known to be non-hazardous. Never use the
instrument with the calibration adapter installed, as this
can cause distorted readings, a potential safety threat.
Note: If the AreaRAE Pro/Plus is equipped with a
gamma sensor, it comes pre-calibrated from the factory
and no calibration is required.
Replacing Battery Packs
The AreaRAE Pro/Plus is powered by a re-
chargeable
To change the battery:
1. Unscrew battery. 2. Lift out. 3. Insert battery.
4. Replace and tighten screws.
USB
Port
Charging
Port
Alarm
Buzzer
4.
WARNINGS
Read Before Operating
CAUTION!
Li-Ion battery pack (PN: W01-3007-000)
User Interface
The user interface consists of the display and three keys,
[Y/+], [MODE] and [N/-]. The LCD displays
information such as monitored threats, real-time
readings, measurement units, alarm type (when in alarm,
including cal. overdue), battery and pump status, datalog
(if on), and radio and connection quality (if available).
RAEMet
Alarm
LEDs
External Filter
And Gas Inlet
2.
[Y/+] key
[MODE] key
External Battery Charger
You can charge a spare Li-Ion battery with the
optional External Li-Ion Battery Charger For
AreaRAE. Attach it to the battery and then plug in
its AC adapter.
Align the Charger's
edges and contact pins
with the battery
Plug in the AC adapter's
barrel and plug the power
cord into an AC source
5.
Antenna
Alarm
LEDs
Display
Charging &
USB ports
(on side)
[N/-] key
3.
Press the Charger into
place, making sure the
contact pins mate
Press down on the back
of the Charger to ensure
that it is securely set
6.

Advertisement

Table of Contents
loading

Summary of Contents for Honeywell AreaRAE Pro

  • Page 1 RAE Systems by Honeywell Display 3775 N. First St. San Jose, CA 95134-1708 USA Note: If the AreaRAE Pro/Plus is equipped with a Charging & External Filter USB ports gamma sensor, it comes pre-calibrated from the factory...
  • Page 2 Align the RAEMet sensor with the socket, using the indexing key and notch as a guide,and press down. Turn Lift the rubber cover on the top of the AreaRAE Pro/Plus the locking ring at the base to tighten until the sensor is firmly seated. Do not turn the top of the sensor.
  • Page 3: Single Sensor Zero Calibration

    Bump Testing and Calibration Setup Fresh Air Calibration Multi-Sensor Span Calibration Manually calibrate using a fixed-flow regulator (flow This procedure determines the zero point for all the Depending on the configuration of your AreaRAE rate between 0.5 and 1.0 liters per minute) and the sensors that require a zero calibration.
  • Page 4 Operation Mode Navigation Menus in Operation Mode are easy to step through by pressing the [N/-] key. Notes:  If the instrument is not equipped with a VOC sensor (PID), or is not equipped with an LEL sensor, then screens for those sensors (VOC Gas Status and LEL Gas Status, respectively) are not shown.
  • Page 5: Single Sensor Span Calibration

    Single Sensor Span Calibration LEL Calibration Procedure Instead of performing a span calibration on 7. A countdown screen appears. You can abort the You must perform a separate zero calibration for the multiple sensors simultaneously, you can select a calibration at any time during the countdown LEL sensor and then perform a span calibration.
  • Page 6: Wireless Operation

    Caution: Refer to the AreaRAE Pro/Plus User’s Guide for safety certification information. Note: If the AreaRAE Pro or AreaRAE Plus is equipped with Wi-Fi, and Wi-Fi is used as the primary radio instead of ISM, then a Wi-Fi access point substitutes for the RAELink3 Host.
  • Page 7: Fcc Part 15 Compliance

    WARNING CAUTION: BEFORE EACH DAY’S USAGE, ONLY THE COMBUSTIBLE GAS DETECTION ANY RAPID UP-SCALE READING FOL- SENSITIVITY OF THE LEL SENSOR MUST BE PORTION OF THIS INSTRUMENT HAS BEEN LOWED BY A DECLINING OR ERRATIC TESTED ON A KNOWN CONCENTRATION OF ASSESSED FOR PERFORMANCE.