ABB MEASUREMENT & ANALYTICS | OPERATING INSTRUCTION
GAA330-M
Emission monitoring system for marine applications
Measurement made easy

Advertisement

Table of Contents
loading

Summary of Contents for ABB GAA330-M

  • Page 1 — ABB MEASUREMENT & ANALYTICS | OPERATING INSTRUCTION GAA330-M Emission monitoring system for marine applications Measurement made easy...
  • Page 3: Table Of Contents

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Contents Preface ..................................5 Safety information ..............................7 Intended application ....................7 Safety information ...................... 7 System overview ..............................10 System description ....................10 System design ......................12 Installation preparation ............................16 Installing the analyzer system ................. 16 Analyzer system: Installation site requirements ..........
  • Page 4 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Operation ................................53 Display and control unit ................... 53 Screen ......................... 54 Status LEDs ........................ 55 Numeric keypad ......................55 Cancel keys ......................... 56 Softkeys ........................56 Password protection ....................58 Menu tree ........................59 Setting the time zone, date and time ..............
  • Page 5: Preface

    • Certificates • Internet You will find information on ABB Analytical products and services on the Internet at "http://www.abb.com/analytical". Service contact If the information in this operating instruction does not cover a particular situation, ABB Service will be pleased to supply additional information as required.
  • Page 6 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Symbols and typefaces in this operating instruction CAUTION identifies safety information to be heeded during analyzer sys- tem operation, in order to avoid risks to the user. NOTE identifies specific information on the operation of the analyzer system as well as on the use of this operating instruction.
  • Page 7: Safety Information

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Safety information Intended application Intended application The analyzer system is designed for continuous measurement of concentra- tions of specific components in gases or vapor. Any other use is not as specified. The specified use also includes taking note of this operating instruction.
  • Page 8 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Safety labels affixed to the analyzer system Observe the safety labels affixed to the analyzer system or to the individual components: Consult documentation! Hot surface! (Temperature > 60 °C) Corrosive material! Risk of electric shock!
  • Page 9 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Risks of a disconnected protective lead The device can be hazardous if the protective lead is interrupted inside or outside the device or if the protective lead is disconnected. Correct operating voltage Be sure the analyzer system voltage setting matches the line voltage before connecting the power supply.
  • Page 10: System Overview

    The analyzer system uses the NDIR photometer AO2000-Uras26 for detecting SO and CO System design The GAA330-M analyzer system is a complete turn-key solution with the following components: Probe and filer unit for proper gas sampling • Heated sample gas line for feeding the sample to the gas analyzer •...
  • Page 11 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Measurement ranges The AO2020-Uras26 gas analyzer is configured and calibrated for the follow- ing measurement ranges: Maximum measurement range for SO is 0 to 500 ppm. • Typical measurement range for SO is 0 to 250 ppm.
  • Page 12: System Design

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C System design System modules Gas analyzer Condensate bottle status indicator Cabinet temperature indicator Bypass flow meter Analyzer flow meter Heated probe temperature controller Measuring point and back-purging pushbuttons Main switch Heated line temperature controller...
  • Page 13 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Gas flow The gas is extracted from the exhaust gas line after the scrubber by a heated sample probe. The heated sampling line is the connection and supplying line between sampling probe and analyzer cabinet. The probe and the sampling line are heated to avoid condensation.
  • Page 14 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Cabinet interior The cabinet contains the sample conditioning components, pump, valves, filters and gas analyzer. It also contains the main switch, the power supply and the fuses for all components, as well as the connection terminals for analog and digital signals.
  • Page 15 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Aqua stop filter The aqua stop filter is the final protection for the analyzer. The aqua stop filter holds back any humidity. Since the aqua stop filter is used as a so called 'police filter', the 1.5 m relaxation distance does not have to be con- sidered.
  • Page 16: Installation Preparation

    OI/GAA330-M-EN REV. C Installation preparation Installing the analyzer system Notes ABB recommends having the analyzer system installed by ABB personnel. • When installing the analyzer system, in addition to this operating in- • struction, comply with the information contained in the drawings set.
  • Page 17: Analyzer System: Installation Site Requirements

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Analyzer system: Installation site requirements CAUTION The analyzer system must not be installed in hazardous locations. Choosing the extraction point The extraction point must be suitable for extracting a representative speci- men flow.
  • Page 18 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Ambient temperature Operation with ventilation fan +5 to +35 °C with cooling unit +5 to +45 °C Storage and transport +2 to +60 °C after draining and drying parts in −25 to +60 °C...
  • Page 19 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Weight of the individual system components Sheet steel cabinet (basic version) Approx. 110 kg Probe tube type 42, heated 8 kg Filter unit, heated, with protective case 20 kg Sample gas line type TBL01, heated...
  • Page 20: Back-Purging Unit: Installation Site And Air Supply Requirements

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Back-purging unit: Installation site and air supply requirements Design of the back-purging unit The back-purging unit consists of a protective cabinet with shut-off valve, 6 bar pressure reduction valve, solenoid valves for back-purging, pressure regulator and 5 l compressed air receiver for effective pressure pulses also with lower airflow rate.
  • Page 21: Items Delivered

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Items delivered Items delivered Quantity Description Standard equipment Analyzer cabinet 1 set System documentation Additional items delivered per order Gas sampling probe tube type 40 (unheated) or type 42 (heated) Filter unit type PFE3 or type PFE2 with heating sleeve and...
  • Page 22: Materials Needed For Installation (Not Supplied)

    Grounding cable ≥ 10 mm Connection between analyzer cabinet and filter unit / sampling probe GAA330-M with filter unit PFE3 and heated probe type 42 Probe type 42 input wiring 2 x 1.5 mm Probe type 42 signal lines (Pt100) 2 x 0.75 mm...
  • Page 23 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Connection between analyzer cabinet and ship control room Analog signal 4 to 20 mA SO 2 x 1 mm (shielded) Analog signal 4 to 20 mA CO 2 x 1 mm (shielded) Alarm and malfunction messages...
  • Page 24: Installation

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Installation Type 40 probe tube and filter unit installation CAUTION The weight of the probe tube with filter unit amounts to approx. 18 to 20 kg! Two persons are needed for transportation and mounting! Before the installation Observe the "Piping plan"...
  • Page 25: Type 42 Probe Tube And Filter Unit Installation

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Type 42 probe tube and filter unit installation CAUTION The weight of the probe tube with filter unit amounts to approx. 28 to 32 kg! Two persons are needed for transportation and mounting! Before the installation Observe the "Piping plan"...
  • Page 26: Pfe2 Filter Unit Installation

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C PFE2 Filter unit installation PFE2 Filter unit: Mounting of probe protective case α Minimum distance x of the mounting flange (wall tube flange) from the wall depending on mounting angle α: 10° 15°...
  • Page 27: Pfe2 Filter Unit Gas Connection

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C PFE2 Filter unit gas connection PFE2 Filter unit: Gas connections (with back-purging)
  • Page 28 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Check valve Solenoid valve for cleaning filter -Y2 Connection for instrument air (max. 6 bar) bulkhead fitting 12 mm Connection for test gas bulkhead fitting 6 mm Connection for Pt100 Heated sample gas line -E2...
  • Page 29: Pfe3 Filter Unit Installation

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C PFE3 Filter unit installation PFE3 Filter unit: Mounting of probe protective case min. 150 25° (dimensions in mm) Minimum distance x of the mounting flange (wall tube flange) from the wall depending on mounting angle α: 10°...
  • Page 30: Pfe3 Filter Unit Gas Connection

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C PFE3 Filter unit gas connection PFE3 Filter unit: Gas connections (with back-purging) Tube test gas, SS 1.4571, 6x1 mm Tube compressed air, Cu, 12x1 mm 1 x M12x1.5 and 1 x M16x1.5 cable connectors 3 x M20x1.5 cable connectors...
  • Page 31: Sample Gas Line Installation

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Sample gas line installation Installing the sample gas line • Observe the "Piping plan" in the drawings set. Connect the sample gas line to the filter unit/gas sampling probe. • Route the sample gas line through the opening provided in the right wall •...
  • Page 32 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Procedures for laying the sample gas line Incorrect Correct Do not lay the heated sample gas Ensure that the hoses do not touch. lines directly side-by-side in an en- Maintain a distance of 25 mm.
  • Page 33 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Incorrect Correct Avoid bundling or laying several Lay several heated sample gas lines heated sample gas lines, so that they separately with a distance of at least touch each other. This results in 2.5 cm and provide adequate ventila-...
  • Page 34: Back-Purging Unit Installation

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Back-purging unit installation Before the installation Observe the "Piping plan" in the drawings set. Installation site The distance between the back-purging unit and the sampling probe must not exceed 5 m (length of the steel-braided compressed-air hoses = 6 m).
  • Page 35: Analyzer Cabinet Installation

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Analyzer cabinet installation Transporting the analyzer cabinet CAUTION The analyzer cabinet weighs approx. 110 kg. A suitable lifting device (crane, block and tackle, lifting truck, etc.) is required for transport, set- ting upright and installation!
  • Page 36: Electrical Connection

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Electrical connection Safety information CAUTION! Please observe the relevant national safety regulations for the construc- tion and operation of electrical installations as well as the following safe- ty instructions. Before connecting the power supply, check that the operating voltage on the rating plate is the same as the mains voltage.
  • Page 37: Type 42 Probe Tube And Filter Unit Electrical Connection

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Type 42 probe tube and filter unit electrical connection Electrical connections 1 2 3 4 5 N N L L Heating Temperature sensor Pt100 Bridge: Power supply 230 V terminals 3 + 4 (shown in the figure)
  • Page 38: Pfe2 Filter Unit Electrical Connection

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C PFE2 Filter unit electrical connection Electrical connections...
  • Page 39 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Terminal block –X1 Supply for heating sleeve of filter unit, 230 VAC, 50 to 60 Hz, 250 W, to be fused by the customer with 6 A, regulation required! Supply for sample gas line, 230 VAC, 50 to 60 Hz, 90 W/m, max.
  • Page 40: Pfe3 Filter Unit Electrical Connection

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C PFE3 Filter unit electrical connection Electrical connections...
  • Page 41: Analyzer Cabinet Electrical Connection

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Analyzer cabinet electrical connection Description of the signal inputs and outputs • Analog outputs: 4 to 20 mA, joint minus pin, electrically isolated, ran- domly groundable, max. DC 30 V, load max. 600 Ω, resolution 16 bit Digital relay outputs: Potential-free contacts (powerless status opened, •...
  • Page 42 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Connection diagrams for the input wiring of the analyzer cabinet...
  • Page 43 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C...
  • Page 44 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C...
  • Page 45: Gas Sampling With Automatic Back-Purging

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Gas sampling with automatic back-purging In general Filter plugging During operation of the analyzer system the dust which is contained in the sample gas will accumulate in the probe filter of the gas sampling system.
  • Page 46: Components For Automatic Back-Purging Procedure

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Components for automatic back-purging procedure Components for automatic back-purging procedure To carry out the automatic back-purging of the filter unit, components are integrated in the analyzer system as follows: the PFE2 filter unit with valves combination for back-purging or •...
  • Page 47: Program Sequence

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Program sequence Program sequence PFE2 Digital -D08 -D08 -D08 -D08 -E05 Status output: Display signal Valve: -Y1.1 -Y2.1 -Y3.1 -Y01 Message Impulse Filter Tube Position "Purge compr. back- back- calibr. back Maint. Step Duration Function...
  • Page 48 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Switch over At first the calibration valve -Y01 is switched over to position "Calibrate". This separates the sample gas conditioning system and the analyzer system from the sampling system and protects it against the back-purging pres- sure.
  • Page 49: Cycle Time

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Cycle time Cycle time duration The cycle time is given as the time interval between two automatic starts of the back-purging procedure. The higher the dust concentration in the sam- ple gas and the higher the sample gas flow, the shorter this time interval must be set, to avoid a blocking of the gas sampling probe filter.
  • Page 50: Post-Purge Time

    Post-purge time = 27 s + 20 s + 1.5 x 5 s + 45 s = 99.5 s Adjustment of the post-purge time To adjust the post-purge time the parameters of the AC500 program must be changed. Please contact ABB service.
  • Page 51: Start-Up And Operation

    NOTE Initial startup of the analyzer system should be performed by trained per- sonnel of the manufacturer or the supplier. ABB recommends having the startup done by ABB personnel. Prior to start-up Check analyzer system seal integrity.
  • Page 52: Warm-Up Phase

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Warm-up phase Warm-up phase The warm-up time is approx. 2 to 4 hours. The warm-up phase can take longer if the analyzer system was not brought to room temperature before the power supply was activated.
  • Page 53: Operation

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Operation Display and control unit Overview Screen with Error Maint Power Menu Line Status LED’s Information Field Numeric Keypad Softkey Line Softkeys Cancel Keys The display and control unit contains: The screen with •...
  • Page 54: Screen

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Screen Screen The backlit graphics has a 320 x 240-pixel resolution. Error Maint Power The screen is divided into three panels Menu line • Information field • Softkey line • The menu line The menu line appears at the upper edge of the screen.
  • Page 55: Status Leds

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Status LEDs Status LEDs The three LEDs next to the screen show the user the gas analyzer’s status. Error Maint Power The green "Power" LED lights when the power supply is on. Power The yellow "Maint"...
  • Page 56: Cancel Keys

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Cancel keys Cancel keys The "Back" and "Meas" keys located under the numeric keypad are designat- Error Maint Power ed as cancel keys. The "Back" key allows the operator to cancel a function or menu item and return to the previous menu level.
  • Page 57 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C The softkeys in measurement mode In the measurement mode, the softkey line contains the softkeys MENU and >>. The softkey Status message also appears if an error occurs. The MENU key is used to call up the main menu and switch to menu mode when in measurement mode.
  • Page 58: Password Protection

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Password protection Elements of password protection Password protection consists of three elements Password level, • User group, • Password. • Password level Each menu item is assigned a password level. Password levels are numbered with 0, 1, 2 and 3.
  • Page 59: Menu Tree

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Menu tree Menu tree For reasons of brevity only the top-level parameters and functions are shown; the menu branches more extensively at most menu items, e.g. into the various measurement components or into the selection and adjustment of values.
  • Page 60: Setting The Time Zone, Date And Time

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Setting the time zone, date and time Menu path MENU → Configure → System → Date/Time Procedure Parameter Explanation Time Zone The time zone can be selected either from the GMT (Green- wich Mean Time) values or from the continent/country/city list.
  • Page 61: Selecting User Interface Language

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Selecting user interface language Menu path MENU → Configure → System → Language Language selection Two user interface languages are factory-configured (per order) in the ana- lyzer system. In the menu item Language, the user can switch between these two languages.
  • Page 62: Maintenance

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Maintenance Safety information CAUTION Only persons familiar with the maintenance of comparable analyzer sys- tems and certified as being capable of such work should work on the sys- tem. Safety labels affixed to the analyzer system...
  • Page 63: Regular Maintenance Tasks

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Regular maintenance tasks NOTE Prior to performing any maintenance works on the analyzer system be sure to activate the "Maintenance Mode" on the "Control Panel" screen thus set- ting the "Maintenance Mode" status signal.
  • Page 64 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Spare parts Description Demand for one year Filter (ceramic) sampling probe Filter (PTFE) inside cabinet Test gas filter Aquastop (upstream the analyzer) Hose for condensate pump Condensate pump Filter material for cabinet fan...
  • Page 65: Analyzer Cabinet: Visual Inspection

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Analyzer cabinet: Visual inspection View of the analyzer system...
  • Page 66 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Visual inspection Device, module Nominal condition Gas analyzer display: Measured values, Green LED "Power" ON status messages Condensate bottle status indicator LED OFF Cabinet temperature indicator < 45 °C Bypass flow meter 120 l/h...
  • Page 67: Calibrating The Analyzer System

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Calibrating the analyzer system CAUTION Only persons familiar with the calibration of comparable analyzer systems are certified as being capable of such work and should work on the sys- tem. NOTE Calibration of the analyzer system has to be performed on board every 6 to 12 months with test gas, in accordance to agreement with class.
  • Page 68 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Adjustment with calibration cell and ambient air The dew point of the inlet air should be similar to the measured gas, for not changing the conditions in the cooler. Start calibration process. The solenoid valve switches to ambient air.
  • Page 69 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Adjustment of the calibration cells with test gas The calibration cells have to be checked in regular intervals with certified test gas. First the analyzer has to be adjusted with test gas. Then the ana- lyzer has to be purged with zero gas and the calibration cell has to be driven into the gas path.
  • Page 70: Checking The Seal Integrity

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Checking the seal integrity When is the seal integrity check needed? Complete seal integrity check of the analyzer system is reserved for certified service personnel. It should be carried out regularly at least every 12 months.
  • Page 71: Replacing The Filter Element In The Filter Unit

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Replacing the filter element in the filter unit CAUTION Prior to performing any maintenance works on the analyzer system be sure to activate the “Maintenance Mode” on the “Control Panel” screen thus setting the “Maintenance Mode” status signal. Be sure to reset this setting after finishing the maintenance work.
  • Page 72 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Filter element T-handle Bridge Detaching disk Locking screw Removal screws Flange O-ring seals Filter element Bridge holding device Casing Casing inner seal (green)
  • Page 73 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Replacing the filter element Turn the T-handle 1 of the filter removal device 1-3 in counter-clockwise direction. This pulls the filter element 8 via the detaching disk 3 out of the casing Turn bridge 2 until it can be pulled off from the bridge holding device 9 through the elongated holes.
  • Page 74 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Turn detaching disk 3 until it can be pulled off from the hexagon screws 5 via the elongated holes. Never loosen or tighten the hexagon screws 5. They have been adjusted at the factory so that the detaching disk 3 can be easily moved.
  • Page 75: Replacing The Cooling Unit / Fan Outlet Filter Mesh

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Replacing the cooling unit / fan outlet filter mesh When is filter mesh replacement needed? The cooling capacity of the cooling unit or the ventilation fan depends upon the cleanness of the filter mesh. It should be replaced if it begins to turn dark.
  • Page 76: Troubleshooting

    Code represents an innovative way of customer’s communication which allows, for instance, improved case-specific support by ABB resulting in an increased availability of analyzer assets. Dynamic QR Code is compatible with the ABB application "my Installed Base" as well as with standard QR code scanner applications. Handling The QR code is selected in the gas analyzer's diagnosis menu and displayed on the gas analyzer's screen.
  • Page 77 The QR code can also be selected in Remote HMI and scanned from the computer screen. Recommended QR code scanner applications ABB recommends the use of the following QR code scanner applications (available free of charge for iOS and Android): "my Installed Base" by ABB...
  • Page 78: Status Messages

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Status messages Information displayed on the ship control system • Calibration analyzer Back-flush active • Collective error • Measuring range selection SO • MR1 0 to 250 ppm • MR2 0 to 500 ppm •...
  • Page 79: Gas Analyzer Problems

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Gas analyzer problems Problem Cause Solution Blinking measurement value Measured signal violates Note: Measurement value > +130 % MRS or readout measurement range limits measurement value < -100% MRS. Additionally, status messages 344 or 345 are generated.
  • Page 80: Analyzer Cabinet Problems

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Analyzer cabinet problems Problem Cause Solution Flow to analyzer not con- Solenoid valve plugged Untie sampling line on both ends of the sole- stant / too low noid valve and carefully clean the valve from...
  • Page 81: Notify Service

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Notify service Who should you contact for further help? Please contact your local service representative. For emergencies, please contact ABB Service, Telephone: +49-(0)180-5-222 580, Telefax: +49-(0)621-381 931 29031, E-mail: automation.service@de.abb.com Before you notify service …...
  • Page 82: Disposal

    If there is no possibility to dispose of the old equipment properly, • ABB service can take care of its pick-up and disposal for a fee. To find your local ABB service contact visit abb.com/contacts or call +49 180 5 222 580.
  • Page 83: Index

    GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. B Index Analyzer cabinet Operation • 53 Visual inspection • 65 Analyzer cabinet electrical connection • 41 Analyzer cabinet installation • 35 Password protection • 58 Analyzer cabinet problems • 80 PFE2 Filter unit electrical connection • 38 Analyzer system PFE2 Filter unit gas connection •...
  • Page 84 GAA330-M EMISSION MONITORING SYSTEM OI/GAA330-M-EN REV. C Notes...
  • Page 86 We reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document.