Page 1
Agent transforms and removes patient identification data before it is securely sent offsite to a Cloud hosted distributed machine data store. The Mini Field Agent can be configured with Aisys CS2 and upgraded Aisys systems. Anesthesia system software version must be 11 (or higher).
Aisys CS2 software version 11 with connectivity and Aisys software version 11 upgrade kit comes with 2 ethernet cables. Software required The Mini Field Agent installation and configuration requires the use of open source WinSCP and PuTTY applications. Download these from: https://winscp.net/eng/download.php...
TCP communication between the anesthesia machine and the Mini Field Agent on port 4444. Network Diagram The Carestation Insights system can support single or multiple operating rooms. One Mini Field Agent device can support up to 10 anesthesia machines. 2094532-001 D 02 2018...
Page 4
MFA WAN Port (required only for configuration and initial setup). Note Ensure the Mini Field Agent is connected to the same facility network to which the anesthesia machines are connected. Note If a firewall or router exists in front of the MFA at the hospital network, then confirm from the network design team or the facility IT team that the entries mentioned in the "Configure Router ACL...
Connect the Mini Field Agent WAN Port 0 to the computer ethernet port, using an ethernet cable. Connect the Mini Field Agent LAN Port 1 to the Facility network switch, using an ethernet cable. Connect the Mini Field Agent to the power supply.
Select OK to exit the Local Area Connection Properties window. 10. Verify that the Mini Field Agent is ON. Note A green LED light beside ON on the Mini Field Agent indicates that the device is switched ON. 11. Click Start. In the Search field, enter: •...
Mini Field Agent. Configuring Facility IP Address To configure the IP address of the Mini Field Agent to connect to the Facility Network, perform the following procedure: Log into WinSCP on the computer connected with the Mini Field Agent.
Page 8
Select SCP in the File protocol drop-down list. In the Host name field, enter: 192.168.101.2 • Verify that the Port number is 22. Modify if required. In the User name field, enter: root • In the Password field, enter: MrootMfa0210 •...
Page 9
12. Save and close the "hosts" file after verification. 13. Navigate to /etc/network 14. Right-click on "interfaces" file and select Edit The "interfaces" file opens as shown in the screenshot: 15. Under # eth0, Change "gateway 192.168.101.1" to "#gateway 192.168.101.1". 16.
Page 10
17. Save the close the "interfaces" file. 18. Reboot the Mini Field Agent as described in "Rebooting Mini Field Agent". 19. Using PuTTY, perform the following: 192.168.101.2 • In the Host Name field, enter: • Ensure port number is 22.
• Configuring Incoming Connection to Mini Field Agent The Mini Field Agent device is required to route the data from the anesthesia machine to the Cloud Gateway. The Mini Field Agent uses a TCP server on port 4444 by default to accept any incoming connections.
Page 12
Right-click on "application.properties" file and select The "application.properties" file opens as shown in the screenshot: Enter the TENANT_ID and CLOUD_URL data based on the details received from GE Carestation Insights Cloud Application team. See "Carestation Insights Cloud information". To edit the timezone: •...
• Navigate to: /root/PredixMachine/etc • Right-click on the com.ge.dspmicro.httpclient.cfg and Edit select The "com.ge.dspmicro.httpclient.cfg" file opens as shown in the below screenshot. Enter the Proxy Host of the facility information to: <Proxy Host> • com.ge.dspmicro.httpclient.proxy.host= Enter the Proxy Port of the facility information to: <Proxy Port>...
The screen looks as follows after the proxy information is added: Configuring DNS Server Settings Note The Installer needs to go through this section and might need to update a file even if there isn’t a DNS server. To configure the DNS Server Settings, perform the following steps: Using WinSCP, navigate to /etc 2094532-001 D 02 2018...
If there are more than one DNS server for a facility, repeat this step for all the DNS Server IP addresses. Save and close the "resolv.conf" file. Rebooting Mini Field Agent To reboot the Mini Field Agent, perform the following steps: Ctrl+T While in WinSCP, press In the Enter command field, enter: •...
Using PuTTY, in the Host Name field, enter: 192.168.101.2 • Verify that the port number is 22. Open Click The Mini Field Agent uses Linux. Log onto the Mini Field Agent using the Linux root login account. root • Enter the Login: MrootMfa0210 •...
MFA. • Using PuTTy, verify that the date on the Mini Field Agent is set to the correct date and time in UTC format, by typing date. An example of date and time in UTC format is Sat May 6 11:14:00 UTC 2017.
• hwclock --show Configuring Anesthesia Machine to Mini Field Agent Perform the following procedure to configure the anesthesia machine to communicate with the SBX Network Client. See Super User mode in the Anesthesia Machine User's Reference Manual for further details.
Page 19
Network Enable Select and set Network Status to Network Config. Select from the right-side button bar. Port 2 In the Network Config. menu, select Enable Set Port 2 to Note Ensure that Port 2 is connected to the Facility network switch, using an ethernet cable.
Page 20
12. Enter the IP Address of the Mini Field Agent. Refer "Facility information" for IP address details. 13. In the Port field, enter: 4444 • 14. Set the Communication Protocol to Sapphire 15. Select Confirm Connectivity Test Start Test 16. Select and then select .
Performing cases with agent, or completed a portion of the system check, sends data to the Mini Field Agent that allows you to confirm the cloud connection of the machine. Complete a case or a portion of the system check to allow you to view the data on the dashboard to confirm.
Page 22
Select the date of the simulated case in the agent cost or checkout dashboard. You can check for the machine serial number (Device ID#) in either of them depending on the method of verification used (ran case or completed checkout). Verify data is present.
Page 23
Keep the anesthesia machines connected. Keep the connection from the Mini Field Agent to the facility network. You can remove the Ethernet cable that is connected from the computer to the Mini Field Agent if you are still connected as it was needed only for configuration purposes.
Page 24
DEFINITIONS TENANT_ID An ID created and provided by GE specific to the customer site. Time Zone (GMT+0:00) Africa/Abidjan (GMT+0:00) Africa/Accra (GMT+1:00) Africa/Algiers (GMT+0:00) Africa/Bissau (GMT+2:00) Africa/Cairo (GMT+0:00) Africa/Casablanca (GMT+1:00) Africa/Ceuta (GMT+0:00) Africa/El_Aaiun (GMT+2:00) Africa/Johannesburg (GMT+3:00) Africa/Juba (GMT+2:00) Africa/Khartoum (GMT+1:00) Africa/Lagos...
MFA to the Cloud network. Verify the following steps: • LAN Port 1 of the Mini Field Agent is connected to the facility network switch, using an ethernet cable. See "Preparing Mini Field Agent for Boot Up".
Configure Router ACL Settings This section is only applicable if a firewall or a router exists in front of the MFA in the facility network. Note Give this information to the facility IT team or the network design team for this configuration. The Network must be configured to allow the below mentioned protocols and ports in the specified directions, irrespective of the variant of router or firewall present on the network.
MFA Mounting Instructions There are two different mounting options for the MFA module: DIN Rail mount or Panel Mount. Panel mounting the MFA requires using the ICEQIMFA-MOUNT panel mount plate. Note Antennae is not needed for this configuration even though an antennae is shown in the below figure.
Page 39
Panel Mounting To panel mount the MFA module, attach the ICEQIMFA-MOUNT panel mount plate to the side of the MFA module using the four M3 screws included with the ICEQIMFA-MOUNT. Attach the panel mount plate in the orientation as displayed in the following figure. The screw size for each panel mount tab is recommended to not exceed M5.