hit counter script
HP E1351A User Manual
HP E1351A User Manual

HP E1351A User Manual

Fet multiplexer
Table of Contents

Advertisement

Quick Links

Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
WARNINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Reader Comment Sheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Chapter 1.Getting Started with the HP E1351A/53A . . . . . . . . . . . . . . . . . . . . . 11
Using This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
FET Multiplexer Module Description . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Multimeter Connection Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Analog Bus Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Tree Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Direct Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Switchbox and Scanning Voltmeter Configurations . . . . . . . . . . . . . . . . . . . 14
Switchbox . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Scanning Voltmeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Digital Bus and Triggering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Programming Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Initial Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Using This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Warnings and Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Multiplexer Card Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Setting the Address Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Selecting the Interrupt Line Number . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Setting the Card ID Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Connecting User Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Adding Signal Conditioning Components/Current Shunts . . . . . . . . . . . . . . . . 23
Connecting Field Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Wiring Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Wiring a Terminal Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Connecting Multimeters and Signal Generators . . . . . . . . . . . . . . . . . . . . . 26
Analog Bus and Digital Bus Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Using This Chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Selecting Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Multiplexer Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Connecting Switchbox Channels to Direct Terminals . . . . . . . . . . . . . . . . . . 32
Comments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Scanning a Range of Switchbox Channels . . . . . . . . . . . . . . . . . . . . . . . . 36
Downloading a Scan List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Scanning a Switchbox without a Downloaded Scan List . . . . . . . . . . . . . . . . . 40
Scanning a Switchbox with a Downloaded Scan List . . . . . . . . . . . . . . . . . . . 41
Scanning Voltmeter Configuration with HP E1326B . . . . . . . . . . . . . . . . . . . 42
HP E1351A/53A FET Multiplexer User's Manual
HP E1351A/53A FET Multiplexer User's Manual Contents
Contents
1

Advertisement

Table of Contents
loading

Summary of Contents for HP E1351A

  • Page 1: Table Of Contents

    Reader Comment Sheet ........9 Chapter 1.Getting Started with the HP E1351A/53A ..... 11 Using This Chapter .
  • Page 2 Measuring Temperature Using Thermocouples (HP E1353A Module only) ....... 43 Setup for Measuring Thermocouple Temperature using an External Multimeter .
  • Page 3 Command Quick Reference ....... . 77 Appendix A.HP E1351A/53A FET Multiplexer Specifications ....79 Appendix B.HP E1351A/53A Register-Based Programming .
  • Page 4 Notes HP E1351A/53A FET Multiplexer User’s Manual Contents...
  • Page 5: Warranty

    For warranty service or repair, this product must be returned to a service facility designated by Hewlett-Packard (HP). Buyer shall pre- pay shipping charges to HP and HP shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to HP from another country.
  • Page 6: Warnings

    DO NOT substitute parts or modify equipment: Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the product. Return the product to a Hewlett-Packard Sales and Service Office for service and repair to ensure that safety features are maintained. HP E1351A, E1353A 16-Channel FET Multiplexer Modules User’s Manual...
  • Page 7: Declaration Of Conformity

    Supplementary Information: The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC and carries the CE-marking accordingly. Tested in a typical configuration in an HP B-Size VXI mainframe. July 20, 1995 Jim White, QA Manager European contact: Your local Hewlett-Packard Sales and Service Office or Hewlett-Packard GmbH, Department HQ-TRE, Herrenberger Straße 130, D-71034 Böblingen, Germany (FAX +49-7031-14-3143).
  • Page 8 Notes HP E1351A, E1353A 16-Channel FET Multiplexer Modules User’s Manual...
  • Page 9: Reader Comment Sheet

    Please fold and tape for mailing Reader Comment Sheet HP E1351A/53A 16-Channel FET Multiplexer Module User’s Manual Edition 4 You can help us improve our manuals by sharing your comments and suggestions. In appreciation of your time, we will enter you in a quarterly drawing for a Hewlett-Packard Palmtop Personal Computer (U.S. government employees cannot participate in the drawing).
  • Page 10 HP E1351A/53A 16-Channel FET Multiplexer Module User’s Manual...
  • Page 11: Chapter 1.Getting Started With The Hp E1351A/53A

    Chapter 1 Getting Started with the HP E1351A/53A Using This Chapter This chapter describes the HP E1351A 16-Channel FET and the HP E1353A 16-Channel Thermocouple FET Multiplexer Modules, and shows how to program the modules using SCPI (Standard Commands for Programmable Instruments) commands.
  • Page 12: Multimeter Connection Points

    Analog Bus The analog bus connector provides a direct connection between multiple multiplexer modules and also between a multiplexer module and HP Connector E1326/E1411 Multimeters. A ribbon cable is used to daisy-chain multiple multiplexer modules together, and to connect a multimeter to the multiplexer modules for a scanning voltmeter.
  • Page 13 Figure 1-1. FET Multiplexer Module Block Diagram Chapter 1 Getting Started with the HP E1351A/53A...
  • Page 14: Switchbox And Scanning Voltmeter Configurations

    The HP-IB interface select code is 7, and the mainframe address is 09. The multiplexer has an instrument address of 24, so its secondary address is 03. The multimeter is HP-IB controlled, and has a primary address of 22.
  • Page 15: Scanning Voltmeter

    (output from multiplexer) signifies advance completed. The HP E1326B Multimeter has a digital bus port on the face plate, and connects to the multiplexer with the digital bus cable (see Figures 2-10 and 2-11).
  • Page 16: Programming Language

    The examples in this manual use the Standard Commands for Programmable Instruments (SCPI) commands, Hewlett-Packard BASIC and an HP 9000 Series 200/300 computer over the HP-IB. Chapter 5 contains information on SCPI command format. Appendix B contains details on the registers for register-based programming.
  • Page 17: Chapter 2.Configuring The Hp E1351A/53A Fet Multiplexer Modules

    STATIC ELECTRICITY. Static electricity is a major cause of component failure. To prevent damage to the electrical components in the multiplexer module, observe anti-static techniques whenever removing a module from the mainframe or whenever working on a module. Chapter 2 Configuring the HP E1351A/53A FET Multiplexer Modules...
  • Page 18: Multiplexer Card Numbers

    Multiplexer Card Numbers HP plug-in modules installed in an HP VXIbus mainframe are treated as independent instruments each having a unique secondary HP-IB address. An instrument may be composed of a single plug-in module or multiple plug-in modules. The card number identifies the module within a switchbox or scanning voltmeter configuration.
  • Page 19: Setting The Address Switch

    Valid address values are Switch from 1 to 255. Refer to the HP 75000 Series B System Installation and Getting Started Guide or the appropriate HP Command Module Manual for addressing information. Otherwise, refer to Figure 2-4 to reset the factory setting.
  • Page 20: Selecting The Interrupt Line Number

    Refer to Figure 2-5 to change the interrupt lines. Valid line numbers are from 1 to 7. Set the interrupt line number to the mainframe interrupt handler number. Figure 2-5. Setting the Interrupt Line Number Configuring the HP E1351A/53A FET Multiplexer Modules Chapter 2...
  • Page 21: Setting The Card Id Switch

    Setting the Card ID Switch The multiplexer module can be an HP E1351A, HP E1352A or HP E1353A depending on which terminal module is installed. With a normal connection, the terminal module will automatically configure the module for the appropriate model. To identify a module without a terminal module, set the Card ID switch as shown in Figure 2-6.
  • Page 22: Connecting User Inputs

    If the terminal module is not desired, Figure 2-7 shows the front panel and the module’s connector pin-out which mates to the terminal module. Figure 2-7. 16-Channel FET Multiplexer Connector Pin-out Configuring the HP E1351A/53A FET Multiplexer Modules Chapter 2...
  • Page 23: Adding Signal Conditioning Components/Current Shunts

    Figure 2-8 shows how to install the appropriate components for the above configurations. In the figure, channel 12 shows a low pass filter configuration, channel 07 an attenuator configuration, and channel 04 a current shunt configuration. Figure 2-8. Signal Conditioning Components/Current Shunts Chapter 2 Configuring the HP E1351A/53A FET Multiplexer Modules...
  • Page 24: Connecting Field Wiring

    Multimeters and signal generators can be connected to the direct terminals or tree terminals on the terminal modules. Figure 2-9 shows the terminal module channel terminals for the 16-Channel FET Multiplexer Module (HP E1351A) and the 16-Channel Thermocouple FET Multiplexer Module (HP E1353A). •...
  • Page 25: Wiring A Terminal Module

    Wiring a Terminal Module The following illustrations show how to connect field wiring to the terminal module. Chapter 2 Configuring the HP E1351A/53A FET Multiplexer Modules...
  • Page 26: Connecting Multimeters And Signal Generators

    Connecting Multimeters and Signal Generators Figure 2-10 shows how to connect the analog bus connector and the digital bus handshake cable between multiple multiplexer modules and the HP E1326B Multimeter. Use the cables shipped with the multiplexer modules to make the connections.
  • Page 27: Analog Bus And Digital Bus Cables

    Analog Bus and Digital Bus Cables Figure 2-11 shows analog bus and digital bus cables for multiplexer-to-multiplexer and multiplexer-to-multimeter connections. Figure 2-11. Analog Bus and Digital Bus Cables Chapter 2 Configuring the HP E1351A/53A FET Multiplexer Modules...
  • Page 28 Notes Configuring the HP E1351A/53A FET Multiplexer Modules Chapter 2...
  • Page 29: Chapter 3.Using The Hp E1351A/53A Fet Multiplexer Modules

    Scanning Voltmeter Configuration with HP E1326B ........Page 42 •...
  • Page 30 CLOSe (@102,208,309) Closes 02 on card 1, 08 on card 2 and 09 on card 3. CLOSe (@103,204) Opens 102 and 208, closes 103 and 204; 309 left closed. Using the HP E1351A/53A FET Multiplexer Modules Chapter 3...
  • Page 31: Multiplexer Commands

    TRIGger[:IMMediate] Channel advance for TRIGger:SOURce BUS | HOLD. source TRIGger:SOURce < > Selects the trigger source as EXTernal, HP-IB BUS, Digital BUS, or IMMediate. < source > = EXT | BUS | DBUS | IMM NOTE: The commands with brackets ([ ]) are implied and are not shown in this chapter.
  • Page 32: Connecting Switchbox Channels To Direct Terminals

    Opens Tree Isolation Switches. CLOSe (@102) Closes channel 02 on card 01. CLOSe (@109) Opens 02 then closes 09. Figure 3-1. Connecting Channels 02 and 09 to their Respective Direct Terminals Using the HP E1351A/53A FET Multiplexer Modules Chapter 3...
  • Page 33: Comments

    If channel 09 is closed, the response will be: 0,0,0,0,0,0,0,0,0,1,0,0,0,0,0,0 NOTE: You must read the response after sending a query command or the switchbox will generate an error. Chapter 3 Using the HP E1351A/53A FET Multiplexer Modules...
  • Page 34: Connecting Switchbox Channels To Tree Terminals For Making Measurements

    SCAN:PORT ABUS Enables tree isolation switches. CLOS (@102) Close channel 02 on card 01. CLOS (@109) Open 02, close 09. Figure 3-2. Closing Channels 02 and 09 for a Voltage Measurement Using the HP E1351A/53A FET Multiplexer Modules Chapter 3...
  • Page 35 Closes channels 02 and 10. 02 is routed to A tree terminal and 10 is routed to B tree terminal. Figure 3-3. Connecting Channels 02 and 10 for 4-Wire Ohms Measurement Chapter 3 Using the HP E1351A/53A FET Multiplexer Modules...
  • Page 36: Scanning A Range Of Switchbox Channels

    You can use any of the modes to advance the channel list. TRIGger requires an external trigger source to be connected TRIGger:SOURce EXTernal to the HP E1300/1301 Mainframe "Event In" port. For computer controlled or HP-IB triggering ( TRIGger *TRG ) use TRIGger:SOURce BUS .
  • Page 37 !Print reading in variable A 110 TRIGGER 70914 !Trigger the switchbox to advance the scan list 120 NEXT I !Increment count 130 END Figure 3-4. Scanning Channels 100 to 215 of a Two-Module Switchbox Chapter 3 Using the HP E1351A/53A FET Multiplexer Modules...
  • Page 38 10, 40 and 60 to the following: 10 OUTPUT 722;"TRIG EXT;OHM" !Set multimeter to 4-wire ohms 40 OUTPUT 70914;"SCAN:MODE FRES"!Closes the appropriate tree isolation switches 60 OUTPUT 70914;"SCAN (@100:107,200:207)" Using the HP E1351A/53A FET Multiplexer Modules Chapter 3...
  • Page 39: Downloading A Scan List

    . The channel closings are controlled by the mainframe CPU. Also, the ARM:COUNt < number > command does not apply to downloaded scan lists. You can use INITiate:CONTinuous ON for continuous scanning with downloaded scan lists. Chapter 3 Using the HP E1351A/53A FET Multiplexer Modules...
  • Page 40: Scanning A Switchbox Without A Downloaded Scan List

    200 NEXT I !Increment counter 210 OUTPUT 70903;"FETC?" !Multimeter retrieves readings from mainframe memory when scan completes 220 ENTER 70903;Rdgs(*) !Put readings in array 230 PRINT Rdgs(*) !Print 240 END Using the HP E1351A/53A FET Multiplexer Modules Chapter 3...
  • Page 41: Scanning A Switchbox With A Downloaded Scan List

    170 OUTPUT 70904;"INIT" !Close first channel 180 OUTPUT 70903;"FETC?" !Multimeter retrieves readings from mainframe memory when scan completes 190 ENTER 70903;Rdgs(*) !Put readings in array 200 PRINT Rdgs(*) !Print 210 END Chapter 3 Using the HP E1351A/53A FET Multiplexer Modules...
  • Page 42: Scanning Voltmeter Configuration With Hp E1326B

    Scanning Voltmeter Configuration with HP E1326B This example shows an HP E1326B Multimeter and an HP E1351A/ E1353A Multiplexer combined into a single VXIbus instrument, a scanning voltmeter. The secondary address for the scanning voltmeter is 03. Both the analog bus connector and the digital bus connector are used. Once the scanning starts, there is no intervention from the mainframe CPU.
  • Page 43: Measuring Temperature Using Thermocouples (Hp E1353A Module Only)

    Measuring Temperature Using Thermocouples (HP E1353A Module only) The thermocouple terminal module automatically configures the HP E1353A FET Multiplexer Module for thermocouple readings. The multimeter measures the resistance of a thermistor on the terminal module to provide a reference temperature. This reference temperature is used to compensate for the change in temperature between the thermocouple (desired measurement) and the terminal temperature (see “Comments”...
  • Page 44: Comments

    Comments Measuring Temperature with the HP El326B/El411B Multimeters. The HP E1326B/E1411B Multimeters can directly measure channels of single or multiple multiplexer modules. The multimeter automatically calculates the correct temperature for the specific thermistor or thermocouple type used. For more information, see the HP E1326B/E1411B Multimeter User’s Manual.
  • Page 45: Chapter 4.Understanding The Hp E1351A/53A Fet Multiplexer Modules

    HP VXIbus mainframes with HP command modules that have "Trig Instruments Out" and "Event In" ports. The computer used in the examples is an HP Series 200/300 used with HP BASIC as the program language. The computer interfaces with the mainframe over HP-IB. Assumed is an: •...
  • Page 46 Figure 4-1. Scanning Commands Understanding the HP E1351A/53A FET Multiplexer Modules Chapter 4...
  • Page 47 As an alternative to OUTP ON, the Channel resource so it is not available for any other Closed on the digital bus also pulses for each instrument in the mainframe. channel closure. Figure 4-2. Trigger Sources Chapter 4 Understanding the HP E1351A/53A FET Multiplexer Modules...
  • Page 48 Example: Scanning This example uses the mainframe "Trig Out" port to synchronize the with External multiplexers to an HP 3457A Digital Multimeter. See the following figure for typical connections. For this example, use the trigger output pulse of the Instruments mainframe "Trig Out"...
  • Page 49 Example: Scanning This example uses the mainframe "Trig Out" and "Event In" ports to Using "Trig Out" and synchronize the multiplexers to an HP 3457A Digital Multimeter. See the following figure for typical connections. "Event In" Ports For this example, use the trigger output pulse of the mainframe’s "Trig Out"...
  • Page 50 A tree terminal 30 OUTPUT 70914;"TRIG:SOUR DBUS"!Sets switchbox to receive external triggers 40 OUTPUT 70914;"SCAN (@100:115)" !Selects the channel list (channels 100 to 115) 50 OUTPUT 70914;"INIT" !Starts scanning cycle 60 END Understanding the HP E1351A/53A FET Multiplexer Modules Chapter 4...
  • Page 51 Example: Scanning In this example the HP E1326B Multimeter is combined with a FET Using Scanning multiplexer to form a scanning voltmeter VXIbus instrument. The multimeter is address 24, and the multiplexer is address 25, so the Voltmeter instrument secondary address is 03. The analog bus connector and the Configuration and digital bus connector are connected between the modules.
  • Page 52: Using The Scan Complete Bit

    Register. When enabled by the STAT:OPER:ENAB 256 command, the Scan Complete Bit will be reported as bit 7 of the Status Register. Use the HP-IB Serial Poll or the IEEE 488.2 Common Command to read the Status *STB? Register. When bit 7 of the status Register is enabled by the...
  • Page 53: Chapter 5.Hp E1351A/53A 16-Channel Fet Multiplexer Command Reference

    SCAN < channel_list > :MODE VOLT [ROUTe:] is the root command, CLOSe SCAN are second-level commands with <channel_list> as a parameter, and MODE is a third-level command with the parameter VOLT Chapter 5 HP E1351A/53A 16-Channel FET Multiplexer Command Reference...
  • Page 54 To set the instrument’s scanning list, you can send either of the following command statements: ROUTe:SCAN (@100:115) SCAN (@100:115) Note You must include a space between the keywords and any parameters. HP E1351A/53A 16-Channel FET Multiplexer Command Reference Chapter 5...
  • Page 55: Linking Commands

    ARM:COUN 1;:TRIG:SOUR EXT SCPI Command Reference This section describes the Standard Commands for Programmable Instruments (SCPI) commands for the multiplexer modules. Commands are listed alphabetically by subsystem and also within each subsystem. Chapter 5 HP E1351A/53A 16-Channel FET Multiplexer Command Reference...
  • Page 56: Abort

    TRIGger:SOURce BUS TRIGger:SOURce HOLD , use ABORt the HP E1301 front panel "Reset Instr "or "Clear Instr "keys to stop the scan. • Stopping Scans Enabled from Front Panel. When a scan is enabled from the HP E1301 front panel execute...
  • Page 57: Arm

    MIN = 1, MAX = 32767 • Comments Related Commands: INITiate[:IMMediate] Example Query Number of Scanning Cycles ARM:COUN 10 Sets 10 scanning cycles. ARM COUN? Query number of scanning cycles; returned value is 10. Chapter 5 HP E1351A/53A 16-Channel FET Multiplexer Command Reference...
  • Page 58: Display

    This command can only be used when the FET multiplexer module is used in a switchbox configuration. It operates with mainframes that have a display, such as the HP 75000 Series B Mainframe (HP Model Number E1301). It also operates with terminals connected to the serial interface port.
  • Page 59: Monitor[:State]

    MONitor:CARD < > | AUTO • Monitor Mode on an HP 1301 Mainframe Display: The following three examples show the types of display you will get: Chan: _ No channels closed on this card. Chan: 12 Channel 12 closed.
  • Page 60: Initiate

    ABORt reset or a power off for downloaded scan lists. • Related Commands: ABORt ARM:COUNt *RST TRIGger TRIGger:SOURce • *RST Condition: INITiate:CONTinuous OFF HP E1351A/53A 16-Channel FET Multiplexer Command Reference Chapter 5...
  • Page 61: Continuous

    An invalid channel list generates an error (see the [ROUTe:]SCAN command). • Stopping Scanning Cycles: See the command. ABORt Example Enabling a Single Scan SCAN (@100:115) Sets channel list. INIT Starts scanning cycle. Chapter 5 HP E1351A/53A 16-Channel FET Multiplexer Command Reference...
  • Page 62: Output

    OUTPut subsystem enables or disables the "Trig Out" port of the HP OUTPut E1300/E1301 Mainframe. Subsystem Syntax OUTPut [:STATe] < state > [:STATe]? [:STATe] state enables/disables the "Trig Out" port on the rear OUTPut[:STATe] < > panel of the HP E1300/E1301 Mainframe.
  • Page 63: [Route:]

    – Closure order for multiple channels with a single command is not guaranteed. • Related Commands: [ROUTe:]OPEN [ROUTe:]CLOSe? [ROUTe:]SCAN:MODE [ROUTe:]SCAN:PORT • *RST Condition: All multiplexer channels are open. Chapter 5 HP E1351A/53A 16-Channel FET Multiplexer Command Reference...
  • Page 64: Close

    Opening order for multiple channels with a single command is not guaranteed. • Open Channels: Opening an already open channel has no effect. • Related Commands: [ROUTe:]CLOSe [ROUTe:]OPEN? • *RST Condition: All multiplexer channels are open. HP E1351A/53A 16-Channel FET Multiplexer Command Reference Chapter 5...
  • Page 65: Open

    CPU during the scan. If an instrument contains mixed modules (FET and Relay Multiplexers), the scan list will not be downloaded. Chapter 5 HP E1351A/53A 16-Channel FET Multiplexer Command Reference...
  • Page 66 See Chapters 2 and 3 for typical user connections to the multiplexer. The computer used in the example is an HP Series 200/300 with HP BASIC as the program language. The computer interfaces with the mainframe over HP-IB.
  • Page 67: Scan:mode

    ), then FRES queries the measurement state. Because 4-wire ohms mode is selected, the query command returns FRES SCAN:MODE FRES Selects the 4-wire ohms scanning mode. SCAN:MODE? Query the scanning mode. Chapter 5 HP E1351A/53A 16-Channel FET Multiplexer Command Reference...
  • Page 68: Scan:port

    This example selects the analog bus port, then queries the state. Because the analog bus port is selected, the query command returns ABUS SCAN:PORT ABUS Selects the analog bus port. SCAN:PORT? Query the port selection. HP E1351A/53A 16-Channel FET Multiplexer Command Reference Chapter 5...
  • Page 69: Settling[:Time]

    Numeric only one channel • Comments Returned value is given in exponential notation. Example Set and Query the SETTling TIME SETT:TIME 16E-6 Sets delay = 16µsec. SETT:TIME? Returns + 1.600000E-005. Chapter 5 HP E1351A/53A 16-Channel FET Multiplexer Command Reference...
  • Page 70: Status

    Reading the Operation Status Register after a Scanning Cycle STAT:OPER? Reads the register value read and print the register value +256 shows bit 8 is set to 1; +0 shows bit 8 is set to 0. HP E1351A/53A 16-Channel FET Multiplexer Command Reference Chapter 5...
  • Page 71: System

    *RST the trigger modes to the power-on states. Example Opening All Channels on Card #1 SYST:CPON 1 Opens all channels on card #1. Chapter 5 HP E1351A/53A 16-Channel FET Multiplexer Command Reference...
  • Page 72: Ctype

    -350, "Too many errors". The least recent error numbers/messages remain in the queue and the most recent are discarded. Example Reading the Error Queue SYST:ERR? Query the error queue. HP E1351A/53A 16-Channel FET Multiplexer Command Reference Chapter 5...
  • Page 73: Trigger

    TRIG:SOUR HOLD Sets trigger source to hold. SCAN (@100:115) Sets channel list. INIT Starts scanning cycle. loop statement Start count loop. TRIG Advances scan to next channel. increment loop Increment count loop. Chapter 5 HP E1351A/53A 16-Channel FET Multiplexer Command Reference...
  • Page 74: Source

    Using External Trigger Inputs: With TRIGger:SOURce EXTernal selected, only one switchbox at a time can use the external trigger input at the HP E1300A/E1301A Mainframe’s "Event In" port. The trigger input is assigned to the first switchbox requesting the external trigger source.
  • Page 75: Source

    Example Scanning Using External Triggers In the following example, the trigger input is applied to the HP E1300A/E1301A Mainframe’s "Event In" port. TRIG:SOUR EXT Sets trigger source to external. SCAN (@100:115) Sets channel list. INIT Starts scanning cycle. trigger externally Advances scan to next channel.
  • Page 76: Ieee 488.2 Common Commands

    16-Channel FET Multiplexer Modules accept. The operation of some of these commands is described in Chapter 4 of this manual. For more information on Common Commands, refer to the HP 75000 Series B Mainframe User’s Manual (HP Model Number E1300/E1301) or the ANSI/IEEE Standard 488.2-1987.
  • Page 77: Command Quick Reference

    When scan is enabled and trigger source is TRIG:SOUR BUS, use the *TRG command to trigger the switchbox to advance the scan. *TST? Self-Test Power on interrupt returns +110. Incorrect MFR ID or model code returns +103. Test pass returns 0. Chapter 5 HP E1351A/53A 16-Channel FET Multiplexer Command Reference...
  • Page 78 Notes HP E1351A/53A 16-Channel FET Multiplexer Command Reference Chapter 5...
  • Page 79: Appendix A.hp E1351A/53A Fet Multiplexer Specifications

    Bandwidth (-3dB): 50Ω source, 1 MΩ | |10pF termination: >200 kHz * Each FET is diode clamped to ±14V and a 220Ω series resistor. This limits the current to the clamp diodes. Appendix A HP E1351A/53A FET Multiplexer Specifications 79...
  • Page 80 Notes 80 HP E1351A/53A FET Multiplexer Specifications Appendix A...
  • Page 81: Appendix B.hp E1351A/53A Register-Based Programming

    Register addresses for register-based devices are located in the upper 25% of VXIbus A16 address space. Every VXIbus device (up to 256 devices) is allocated a 32 word (64 byte) block of addresses. The HP E1351A/E1353A multiplexers use eight two-byte registers of the 64 addresses (32 word registers) allocated.
  • Page 82 Figure B-1. Multiplexer Registers within A16 Address Space Figure B-2. Registers within HP E1301/E1406 A16 Address Space 82 HP E1351A/53A Register-Based Programming Appendix B...
  • Page 83: The Base Address

    This address consists of a base address plus a register offset. The base address used programming depends on whether the A16 address space is outside or inside the HP E1405A/E1406A Command Module or HP E1300A/E1301A Mainframe. A16 Address Space When the command module or mainframe is not part of your VXIbus system (Figure B-1), the multiplexer’s base address is computed as:*...
  • Page 84: Register Offset

    Status/Control Register (base + 04 • Scan Control Register (base + 06 • Scan Channel Delay Register (base + 08 Manufacturer ID Register base + 00 Write Undefined Read Device Address Manufacturer ID Class Space 84 HP E1351A/53A Register-Based Programming Appendix B...
  • Page 85 + 0A Write VLD* A_D* B_D* Read Undefined Direct Channel Configuration Register base + 0C Write Undefined Read Undefined Direct Control Register base + 0E Write Undefined VLD* A_D* B_D* Read Undefined Appendix B HP E1351A/53A Register-Based Programming 85...
  • Page 86: Register Bit Descriptions

    Write Undefined Read Model Code Model Code Model Code. The following model codes identify the HP E1351A FET Multiplexer and HP E1353A Thermocouple FET Multiplexer: F F 1 0 HP E1351A 16-Channel FET Multiplexer F F 1 2 HP E1353A 16-Channel Thermocouple FET Multiplexer...
  • Page 87: Status/Control Register

    BSY* is unasserted at the beginning of the Channel Closed pulse. If a trigger occurs during the Channel Closed pulse, the pulse is terminated and the channel is advanced. Operates only with downloaded scan list (DIR is zero). Appendix B HP E1351A/53A Register-Based Programming 87...
  • Page 88: Scan Control Register

    WRITE register. base + 08 Write Undefined Read Undefined D3 - D0 D0 is the least significant bit, D3 is the most significant. These bits designate a number between 0 and 15 for 88 HP E1351A/53A Register-Based Programming Appendix B...
  • Page 89: Scan Channel Configuration Register

    Read Undefined D3 - D0 These data bits set the channel number with a range of 0 - 15. Bit D0 is the least significant bit, and bit D3 is the most significant. Appendix B HP E1351A/53A Register-Based Programming 89...
  • Page 90: Direct Control Register

    HP E1326B Multimeter User’s Manual. ABORt ABORt does not have an equivalent in register-based programming. This command is executed from the front panel or over HP-IB. ARM:COUNt ARM:COUNt command does not have an equivalent in register-based programming. To scan for a specified number of cycles, use the interrupt generated at the end of the scan list to advance through a program loop for the set number of cycles.
  • Page 91 100. VLD* false (1), XXXX to D3-D0 Reserves position in scan list for channel 101. VLD* true (0), 0000 to D3-D0 Sets channel 200. VLD* true (0), 0001 to D3-D0 Sets channel 201. Appendix B HP E1351A/53A Register-Based Programming 91...
  • Page 92 These commands both write to the TRG INT bit when they are active and a trigger occurs. is not implemented at the register level. It is in TRIGger:SOURce HOLD the normal state when all trigger sources are disabled. 92 HP E1351A/53A Register-Based Programming Appendix B...
  • Page 93: Register-Based Programming For Maximum Speed

    E1326B microprocessor as you can easily write commands faster than the processor can execute them. 2. In order to program the HP E1326B Multimeter to take readings with the HP E1351A FET Multiplexer at its highest speed, do the commands from Table B-1 on page 96.
  • Page 94 Configuration Register. This register loads up a FIFO that will later rotate each time the HP E1326B issues a VM complete on the digital bus. If you have more than one FET multiplexer, you must load up the FIFOs on each one.
  • Page 95 5. The voltmeter is now taking data which is being stored in the HP E1326B hardware FIFO. With the 10 µsec aperture set, 512 16-bit readings will collect here.
  • Page 96: Multimeter Command And Parameter Opcodes

    — - —- Trigger Arm Un-armed —- — - Armed Trigger Immediate 13 — - — - —- Send Error — - — - —- — - Table B-1 Continued on Next Page 96 HP E1351A/53A Register-Based Programming Appendix B...
  • Page 97 (e.g. Trigger Immediate). Subsequent measurements in the burst occur when Software Sample is written to the Command register. (7) Up to seven VXIbus TTLTrg trigger lines can be selected. 01 selects line 0, FF selects all lines. Figure B-3. Configuring the Multimeter Appendix B HP E1351A/53A Register-Based Programming 97...
  • Page 98: Multimeter Register-Based Programming Error Codes

    > +- 40V between HI and LO or HI and COMMON with the range <= 8V > +- 40V between LO and COMMON on any range Under this condition, the multimeter disconnects itself from the input. A new range must be specified to restore operation. 98 HP E1351A/53A Register-Based Programming Appendix B...
  • Page 99: Appendix C.hp E1351A/53A Fet Multiplexer Error Messages

    HP E1351A/53A FET Multiplexer Error Messages Table C-1 lists the error messages associated with the multiplexer modules programmed with SCPI commands. See the HP E1406A Command Module User’s Manual for complete information on error messages. Table C-1. 16-Channel FET Multiplexer Error Messages...
  • Page 100 Notes 100 HP E1351A/53A FET Multiplexer Error Messages Appendix C...
  • Page 101 Certification, 5 Analog Bus Channel cable, 26 - 27 address, 30 cables, 42 advance triggering, 12, 15 connector, 12, 26 - 27 advancing, 36 port, 68 closing, 16, 29, 33 - 35, 63 HP E1351A/53A FET Multiplexer User’s Manual Index...
  • Page 102 Direct Channel Configuration Register, 85, 89 *STB?, 52, 76 Direct Control Register, 85, 90 *TRG, 76 - 77 Direct Terminals *TST?, 76 - 77 connecting switchbox channels, 32 - 33 *WAI, 76 description of, 12 HP E1351A/53A FET Multiplexer User’s Manual Index...
  • Page 103 Downloading Scan List, 39 - 41, 65 scan rate, 36 scanning, 93 switching, 11 - 12 temperature measurements, 44 Errors HP 3457A Digital Multimeter, 48 - 50, 66 codes, register-based, 98 HP E1326B Multimeter, 51 messages, 72, 99 - 100 HP-IB numbers, 72, 99 - 100...
  • Page 104 41 high-speed scanning, 93 scanning with external instruments, 48 measuring thermocouple temperature, 43 - 44 scanning without a scan list, 40 primary address, 14 verify initial operation, 16 HP E1351A/53A FET Multiplexer User’s Manual Index...
  • Page 105 39, 65 within A16 address, 82 scanning with downloading, 41 WRITE, 84 Ribbon Cable, 12 scanning without downloading, 40 Scanning Voltmeter ROUTe: Subsystem, 63 - 69 analog bus connector, 12 ROUTe:CLOSe, 63, 91 HP E1351A/53A FET Multiplexer User’s Manual Index...
  • Page 106 44 Standard Commands for Programmable Instruments resistance, converting to a temperature, 44 See SCPI Commands temperature, converting to a voltage, 44 Static Electricity, 17 Thermocouple, 43 Status Register, 52, 70 compensated measurements, 44 HP E1351A/53A FET Multiplexer User’s Manual Index...
  • Page 107 15 - 16, 50 - 51, 74 external inputs, 74 source, 12, 45, 47, 74 *TST?, 76 - 77 Two-wire Ohms Measurements, 38, 67 Understanding the HP E1351A/53A, 45 - 52 Using bus triggers, 74 - 75 external trigger inputs, 74 external triggers, 75...

This manual is also suitable for:

E1353a

Table of Contents