GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
Section 2-2
Console Requirements
2-2-1
Unit Environmental Requirements
Table 2-9
Requirement
Operational
Storage
Transport
CAUTION: If the system has been in storage or has been transported, please see the acclimation
requirements before powering
on page 3-2.
2-2-2
Cooling Requirements
The cooling requirement for the Vivid 3N Pro/Expert ultrasound unit environment is
3500 BTU/hr. This figure does not include the cooling required for lights, people, or other equipment in
the room.
Note:
Each person in the room places an additional 300 BTU/hr demand on the environmental cooling.
2-2-3
Lighting Requirements
For system installation, updates and repairs, bright lighting is required. However, operator and patient
comfort may be optimized if the room lighting is subdued and indirect when a scan is being performed.
Therefore, a combination lighting system (dim/bright) is recommended. Keep in mind that lighting
controls and dimmers can be a source of EMI which could degrade image quality. These controls should
be selected to minimize possible interference.
2-2-4
Time and Manpower Requirements
Site preparation takes time. Begin pre-installation checks as soon as possible to allow sufficient time to
make any required changes. If possible, begin these checks as many as six weeks before system
delivery.
CAUTION: At least two people must be available to deliver and unpack the Vivid 3N Pro/Expert
ultrasound unit. Attempts to move the unit considerable distances (or on an incline) by one person
alone, could result in personal injury, and/or damage to the system.
2-2
04
EVISION
Environmental Requirements
Temperature
o
10 — 40
C
o
-20 — 60
C
o
-20 — 60
C
ON
Section 2-2 - Console Requirements
Relative Humidity
(non-condensing)
o
50 — 70%
(50 — 104
F)
o
10 — 95%
(-4 — 140
F)
o
10 — 95%
(-4 — 140
F)
and/or using the system.
V
3N P
/E
IVID
RO
XPERT
Air Pressure
700 — 1060 hPa
700 — 1060 hPa
700— 1060 hPa
Refer to the Installation Warnings section
S
M
ERVICE
ANUAL